Supplemental oxygen must not be used while smoking or in the presence of an open 

flame. 

 

Always make sure that the device is turned on and airflow generated before the oxygen 

supply is turned on. Always turn the oxygen supply off before the device is turned off, so 

that unused oxygen does not accumulate within the device enclosure and create a risk of 

fire. 

 

Do not perform any maintenance tasks while the device is in operation. 

 

The device should not be used adjacent to or stacked with other equipment. If adjacent or 

stacked use is necessary, the device should be observed to verify normal operation in the 

configuration in which it will be used. 

 

The use of accessories other than those specified for the device is not recommended. They 

may result in increased emissions or decreased immunity of the device. 

  CAUTION 

 

Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the 

effectiveness of the treatment and/or damage the device. 

 

Use only vented masks recommended by ResMed or by the prescribing doctor with this 

device. Fitting the mask without the device blowing air can result in rebreathing of exhaled 

air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow 

of the fresh air into the mask. 

 

Be careful not to place the device where it can be bumped or where someone is likely to 

trip over the power cord. 

 

Blocking the air tubing and/or air inlet of the device while in operation could lead to 

overheating of the device. 

 

Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding) 

that could block the air inlet or cover the power supply unit. 

 

Do not place the device on its side as water might get into the device. 

 

Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is 

correctly set up. 

 

Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or 

antibacterial soaps or scented oils to clean the device, the water tub or air tubing. These 

solutions may cause damage or affect the humidifier performance and reduce the life of 

the products. 

 

If you use the humidifier, always place the device on a level surface lower than your head 

to prevent the mask and air tubing from filling with water. 

 

Leave the water tub to cool for ten minutes before handling to allow the water to cool and 

to make sure that the water tub is not too hot to touch. 

 

Make sure that the water tub is empty before transporting the device. 

 

 

English 

15

 

Summary of Contents for AirCurve10 ASV

Page 1: ...User guide English User guide...

Page 2: ...with chronic symptomatic heart failure NYHA 2 4 with reduced left ventricular ejection fraction LVEF 45 and moderate to severe predominant central sleep apnea Positive airway pressure therapy may be...

Page 3: ...se with the device including Air tubing heated and non heated ClimateLineAir SlimLine Standard Water tub Standard water tub cleanable water tub can be disinfected Filter Hypoallergenic filter standard...

Page 4: ...select an option Turn to adjust a selected option and press to save your change Home button Press to return to the Home screen Different icons may be displayed on the screen at different times includi...

Page 5: ...to the power outlet 3 Connect the air tubing firmly to the air outlet located on the rear of the device 4 Open the water tub and fill it with distilled water up to the maximum water level mark Do not...

Page 6: ...that adjusts the screen brightness based on the light in the room Stopping therapy 1 Remove your mask 2 Press Start Stop or if SmartStart is enabled therapy will stop automatically after a few seconds...

Page 7: ...e you can personalize your options Ramp Time Designed to make the beginning of therapy more comfortable Ramp Time is the period during which the pressure increases from a low start pressure to the pre...

Page 8: ...air 3 Adjust the mask mask cushion and headgear until you get a Good result To stop Mask Fit press the dial or Start Stop If you are unable to get a good mask seal talk to your care provider More opti...

Page 9: ...ining water 3 Hold the cuff of the air tubing and gently pull it away from the device 4 Hold both the cuff of the air tubing and the swivel of the mask then gently pull apart Cleaning You should clean...

Page 10: ...water tub and air tubing are dry you can reassemble the parts 1 Connect the air tubing firmly to the air outlet located on the rear of the device 2 Open the water tub and fill it with distilled room t...

Page 11: ...and send it back to your care provider For more information on the SD card refer to the SD card protective folder provided with your device Note The SD card should not be used for any other purpose Tr...

Page 12: ...troubleshooting Problem possible cause Solution Air is leaking from around my mask Mask may be fitted incorrectly Make sure your mask is fitted correctly See your mask user guide for fitting instructi...

Page 13: ...d where there is coverage ie on your bedside table not in a drawer or on the floor The Wireless signal strength icon indicates good coverage when all bars are displayed and poor coverage when fewer ba...

Page 14: ...e lock read only position Move the switch on the SD Card from the lock position to the unlock position and then re insert it System fault refer to user guide Error 004 Device may have been left in a h...

Page 15: ...and plug are in good condition and the equipment is not damaged Keep the power cord away from hot surfaces If you notice any unexplained changes in the performance of the device if it is making unusu...

Page 16: ...aled air Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask Be careful not to place the device where it can be bumped or where someone...

Page 17: ...he electromagnetic emissions and immunity of this ResMed device can be found on www resmed com on the Products page under Service and Support IEC 60601 1 2005 classification Class II double insulation...

Page 18: ...3G USA and Canada only The AirCurve 10 device complies with Industry Canada license exempt RSS standards The AirCurve 10 device should be used at a minimum distance of 0 8 2 cm from the body during op...

Page 19: ...lly conductive or antistatic air tubing The temperature and relative humidity settings displayed are not measured values Displayed values Value Range Display resolution Pressure sensor at air outlet M...

Page 20: ...0 5 0 8 0 8 12 0 5 0 5 0 5 0 5 0 8 0 8 16 0 5 0 5 0 5 0 5 0 8 0 8 20 0 5 0 5 0 5 0 5 0 8 0 8 25 0 4 0 3 0 6 0 5 0 8 0 8 Symbols The following symbols may appear on the product or packaging Read instr...

Page 21: ...your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period Mask systems including mask frame cushion h...

Page 22: ...incidental or consequential damages claimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or c...

Page 23: ...96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip AirCurve Air10 ClimateLine HumidAir and...

Reviews: