Problem/possible cause 

Solution 

Air pressure in my mask seems too low (it feels like I am not getting enough air) 

Ramp may be in progress. 

Wait for air pressure to build up or turn Ramp Time off. 

My screen is black 

Backlight on the screen may have turned off. It turns off 

automatically after a short period of time. 

Press Home or the dial to turn it back on. 

Power may not be connected. 

Connect the power supply and make sure the plug is fully 

inserted. 

I have stopped therapy, but the device is still blowing air 

Device is cooling down. 

Device blows a small amount of air in order to avoid 
condensation in the air tubing. It will stop automatically after 

20 minutes. 

My water tub is leaking 

Water tub may not be assembled correctly. 

Check for damage and reassemble the water tub correctly. 

Water tub may be damaged or cracked. 

Contact your care provider for a replacement. 

My therapy data has not been sent to my care provider 

Wireless coverage may be poor. 

Make sure that the device is placed where there is coverage 
(ie, on your bedside table, not in a drawer or on the floor). 
The Wireless signal strength icon 

  indicates good 

coverage when all bars are displayed, and poor coverage 

when fewer bars are displayed. 

The No wireless connection icon 

is displayed on the 

top right of the screen. No wireless network available. 

Make sure that the device is placed where there is coverage 

(ie, on your bedside table, not in a drawer or on the floor). 
If instructed to do so, send the SD card to your care provider. 

The SD card also contains your therapy data. 

Device may be in Airplane Mode. 

Turn off Airplane Mode, see Traveling by plane. 

Data transfer is not enabled for your device. 

Contact your care provider to enable the data transfer 

service. 

My screen and buttons are flashing 

Software upgrade is in progress. 

Software upgrade takes approximately 10 minutes to 

complete. 

Device messages 

Device message/possible cause 

Solution 

High leak detected, check your water tub, tub seal or side cover 

Water tub may not be inserted properly. 

Make sure the water tub is correctly inserted. 

Water tub seal may not be inserted properly. 

Open the water tub and make sure that the seal is correctly 

inserted. 

12

Summary of Contents for AirCurve10 ASV

Page 1: ...User guide English User guide...

Page 2: ...with chronic symptomatic heart failure NYHA 2 4 with reduced left ventricular ejection fraction LVEF 45 and moderate to severe predominant central sleep apnea Positive airway pressure therapy may be...

Page 3: ...se with the device including Air tubing heated and non heated ClimateLineAir SlimLine Standard Water tub Standard water tub cleanable water tub can be disinfected Filter Hypoallergenic filter standard...

Page 4: ...select an option Turn to adjust a selected option and press to save your change Home button Press to return to the Home screen Different icons may be displayed on the screen at different times includi...

Page 5: ...to the power outlet 3 Connect the air tubing firmly to the air outlet located on the rear of the device 4 Open the water tub and fill it with distilled water up to the maximum water level mark Do not...

Page 6: ...that adjusts the screen brightness based on the light in the room Stopping therapy 1 Remove your mask 2 Press Start Stop or if SmartStart is enabled therapy will stop automatically after a few seconds...

Page 7: ...e you can personalize your options Ramp Time Designed to make the beginning of therapy more comfortable Ramp Time is the period during which the pressure increases from a low start pressure to the pre...

Page 8: ...air 3 Adjust the mask mask cushion and headgear until you get a Good result To stop Mask Fit press the dial or Start Stop If you are unable to get a good mask seal talk to your care provider More opti...

Page 9: ...ining water 3 Hold the cuff of the air tubing and gently pull it away from the device 4 Hold both the cuff of the air tubing and the swivel of the mask then gently pull apart Cleaning You should clean...

Page 10: ...water tub and air tubing are dry you can reassemble the parts 1 Connect the air tubing firmly to the air outlet located on the rear of the device 2 Open the water tub and fill it with distilled room t...

Page 11: ...and send it back to your care provider For more information on the SD card refer to the SD card protective folder provided with your device Note The SD card should not be used for any other purpose Tr...

Page 12: ...troubleshooting Problem possible cause Solution Air is leaking from around my mask Mask may be fitted incorrectly Make sure your mask is fitted correctly See your mask user guide for fitting instructi...

Page 13: ...d where there is coverage ie on your bedside table not in a drawer or on the floor The Wireless signal strength icon indicates good coverage when all bars are displayed and poor coverage when fewer ba...

Page 14: ...e lock read only position Move the switch on the SD Card from the lock position to the unlock position and then re insert it System fault refer to user guide Error 004 Device may have been left in a h...

Page 15: ...and plug are in good condition and the equipment is not damaged Keep the power cord away from hot surfaces If you notice any unexplained changes in the performance of the device if it is making unusu...

Page 16: ...aled air Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask Be careful not to place the device where it can be bumped or where someone...

Page 17: ...he electromagnetic emissions and immunity of this ResMed device can be found on www resmed com on the Products page under Service and Support IEC 60601 1 2005 classification Class II double insulation...

Page 18: ...3G USA and Canada only The AirCurve 10 device complies with Industry Canada license exempt RSS standards The AirCurve 10 device should be used at a minimum distance of 0 8 2 cm from the body during op...

Page 19: ...lly conductive or antistatic air tubing The temperature and relative humidity settings displayed are not measured values Displayed values Value Range Display resolution Pressure sensor at air outlet M...

Page 20: ...0 5 0 8 0 8 12 0 5 0 5 0 5 0 5 0 8 0 8 16 0 5 0 5 0 5 0 5 0 8 0 8 20 0 5 0 5 0 5 0 5 0 8 0 8 25 0 4 0 3 0 6 0 5 0 8 0 8 Symbols The following symbols may appear on the product or packaging Read instr...

Page 21: ...your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below Product Warranty period Mask systems including mask frame cushion h...

Page 22: ...incidental or consequential damages claimed to have resulted from the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or c...

Page 23: ...96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip AirCurve Air10 ClimateLine HumidAir and...

Reviews: