No head movement in automatic mode
Possible cause
Solution
Head controller in manual mode
Change to automatic mode.
Power loss
Check mains cable connections and integrity.
Check power block is supplying 24 V.
Check CMM emergency stop condition - power may have been
removed by CMM.
Cable / connection fault
Check connection and integrity of cabling between head and
controller.
Communications failure
Check RS232 baud rate.
Incorrect command set selected
Check correct command set is selected.
PICS STOP is asserted by other system component
Clear the fault asserting PICS STOP.
No head movement in manual mode
Possible cause
Solution
Head controller in automatic mode.
Change to manual mode.
Power loss
Check mains cable connections and integrity.
Check power block is supplying 24 V.
Check CMM emergency stop condition - power may have been
removed by CMM.
Cable / connection fault
Check connection and integrity of cabling between head and
controller.
PH10M-iQ PLUS installation and user's guide
www.renishaw.com
Issued 07 2021
31