11
SPECIFICATION
dimensions mm (in)
PROBE REPEATABILITY
Maximum 2 Sigma (2
σσσσσ
) Value
Repeatability of 1,0µm (0.00004in) is valid for
test velocity of 480mm/min (1.57ft/min) at
stylus tip, using stylus 50mm (1.97in) long.
SENSE DIRECTIONS and
STYLUS TRIGGER FORCE
Set at factory using stylus 50mm (1.97in) long.
X and Y trigger forces vary around the stylus
seating.
X and Y direction - lowest force 0,75N
75gf
(2.64oz)
X and Y direction - highest force 1,4N
140gf
(4.92oz)
Z direction force
4,9N
490gf
(17.28oz)
PROBE SEALING
IPX8 (BS 5490, IEC 529)
100kPa
(1 atmosphere).
SYSTEM DELAYS
The time delay from the probe stylus contacting
the workpiece, and the MI16 machine interface
providing an output signal to the CNC controller
is 12ms ±10µs.
STYLUS OVERTRAVEL LIMITS
21,5mm
(0.85in)
Z
X/Y
17.5° 17.5°
Stylus
length
50mm
(1.96in)
100mm
(3.93in)
X - Y
21,5mm
(0.84in)
36,5mm
(1.43in)
Z
8mm
(0.31in)
8mm
(0.31in)
TEMPERATURE
PROBE/RMP3
RMM2
MI16 INTERFACE
OM16
ENVIRONMENT
Region
USA
RADIO FREQUENCIES and TRANSMISSION RANGE
.
Nominal
transmission
frequency
433 MHz
Number of
transmission
channels
69
Channel
separation
25 KHz
Transmission
range
3m (9.8ft)
Storage
-20°C to 60°C
(-4°F to 140°F)
Operating
0°C to 40°C
(32°F to 104°F)