3.3 Liability Exclusion
Renfert shall be absolved from all claims for damages
or warranty if:
• The product is employed for any purposes
other than those cited in the operating
instructions;
• The product is altered in any way other than
those alterations described in the operating
instructions;
• The product is repaired by other than an
authorized facility or if any but Renfert OEM
parts are employed;
• The product continues to be employed, despite
obvious safety faults or damage;
• The product is subjected to mechanical
impacts or is dropped.
4. Installation
4.1 Wall Mounting
(with the included mounting set)
Please have the following on hand:
• pencil,
• drilling template,
• Philips head screwdriver,
• power drill,
• 8 mm [0.3152 inch] drill bit - suitable for penetrat-
ing the intended wall material.
Inspect the wall on which you intend mount-
ing the unit to ensure that it is capable of
bearing the unit’s weight.
1. Determine the optimum working height (Figure 1).
2. Align the drilling template and mark the drill holes
(Figure 2).
3. Drill the required holes (using the 8 mm [0.3152
inch] drill bit
>>> to a depth of at least 55 mm [2.167 inches]) +
insert the dowels (Figure 3).
4. Screw the top screws into the dowels fi rst
>>> allowing the screw heads to protrude
9-10 mm [0.3546 – 0.394 inches] (Figure 4).
5. Mount the bottom mounting bracket (Figure 5).
6. Hang the unit on the wall (Figure 6).
7. Secure the unit with the knurled nut
>>> Securely tighten the knurled nut (Figure 7).
4.2 Benchtop Units
1. Position the stand on a level surface (Figure 8).
2. Hang the unit in the stand (Figure 9).
3. Secure the unit with the knurled nut
>>> Securely tighten the knurled nut (Figure 10).
The stand is not part of the standard delivery and
must be ordered as an optional accessory (refer to
the "Accessories" section).
5. Operating Elements
(Figure 11 + 12)
A Vacuum display
B Mixing parameters display (e.g., speed, mixing
time, etc.), refer to Table 5.1
C Program number indicator
D Programming key, "P"
E Parameter setting keys
F Control knob
G On/Off switch
H Power cord
J Unit fuse
K Motor fuse
L Bowl detection pin
M Vacuum fi lter
5.1 Key Symbols /
Adjustment Ranges
Mixing
parameters
Key
symbol
Adjustment
range
Factory
setting
Premix,
plaster
0:00 / 0:15 /
0:20 / 0:25
0:00
Premix,
investment
0:00 / 0:15 /
0:20 / 0:25
0:00
Pre-vacuum
0:00 - 1:00 min.
0:00
Speed
150 – 600 1/min. 350
Mixing time
0:00 - 5:00 min.
0:30
Interval
(*) (direction
change)
0:00 - 0:30 min.
0:00
Post-
vacuum
0 - 1 min.
0:00
(*) Interval:
The time period after which the current direction of mixing is
reversed. This only applies if the mixing time is longer than the
interval.