5.2 Troubleshooting When Using Emulator Debugger M3T-PD308M
When M3T-PD308M has started up normally, however, errors occurred while using it, remedy the
troubles referring Table 5.4.
Table 5.4 Error message and how to remedy it when using M3T-PD308M
Error message
Checkpoint
Target clock is stopped
Target MCU is unable to reset.
Target MCU is reset state.
Please reset target systems.
Target MCU is HOLD state.
Target MCU is not given power.
Connection to the
target system
- If an external clock is being supplied, check
the target system's oscillation circuit works
properly. Also, if using the sub-clock, see "3.4
Using the P8
7
/X
CIN
Pin as Sub-Clock Input"
(page 28).
- If the external clock supply is set, change to
internal supply. Also, if using the sub-clock,
see "3.4 Using the P8
7
/X
CIN
Pin as Sub-Clock
Input" (page 28).
- If an external clock is being supplied, check
the target system's oscillation circuit works
properly. Also, if using the sub-clock, see "3.4
Using the P8
7
/X
CIN
Pin as Sub-Clock Input"
(page 28).
- If the external clock supply is set, change to
internal supply. Also, if using the sub-clock,
see "3.4 Using the P87/X
CIN
Pin as Sub-Clock
Input" (page 28).
- The target MCU has been reset. Cancel the
reset.
- If an external clock is being supplied, check
the target system's oscillation circuit works
properly. Also, if using the sub-clock, see "3.4
Using the P8
7
/X
CIN
Pin as Sub-Clock Input"
(page 28).
- The MCU may be in the stop or wait modes.
Either reset the MCU or cancel the mode with
an interrupt.
- If an external clock is being supplied, check
the target system's oscillation circuit works
properly. Also, if using the sub-clock, see "3.4
Using the P8
7
/X
CIN
Pin as Sub-Clock Input"
(page 28).
- The MCU may be in the stop or wait modes.
Either reset the MCU or cancel the mode with
an interrupt.
- Check the power supply and GND are properly
connected to the target system.
Connected
Connected
Not connected
Connected
Not connected
Connected
Not connected
Connected