Maintenance
26 / 44
TARAtec 9/10
7.2
Changing the electrolyte and membrane cap
Unscrew the membrane cap.
Using tweezers, remove the G-holder from the membrane cap.
Empty the electrolyte out of the membrane cap.
Rinse the membrane cap with mains water.
Rinse the electrode finger with mains water.
Shake the sensor body dry several times (see Fig. 12, p. 25).
This empties the pressure compensation opening [1].
Fig. 12: Shaking the sensor body dry
1
Pressure compensation opening