Information
on this product
6 / 44
TARAtec 9/10
2
Information on this product
2.1
Product description
The TARAtec 9/10 product range consists of sensors with covering
membranes.
They are 2-electrode systems for measuring the concentrations in water
of the disinfectants chlorine
1
, chlorine dioxide, ozone, peracetic acid or
hydrogen peroxide. In comparison to the product range TARAtec 7 the
TARAtec 9/10 sensors are characterised by improved resistance to
media for reducing surface tension (tensides). In addition, TARAtec 9
sensors can be used when the water being measured is at high
temperatures. The area of application of these sensors extends to almost
all water qualities.
The sensors are not suitable for checking the absence of chlorine,
chlorine dioxide, ozone, peracetic acid or hydrogen peroxide.
A complete measuring and/or control system normally consists of the
following components:
Sensor
Electrical leads and connectors
Flow chambers and connections
Measuring and control device
Dosing equipment
Analytical equipment
These operating instructions relate exclusively to the sensor.
Comply with the operating instructions for the peripheral
devices.
1
Free chlorine (dependent on the pH value)