SPECIFICATIONS:

 

Catalog 

13940013

Reading Scale -  

DEF/AdBlue™  = Diesel exhaust fluid, % urea by  

 

 

w/w (weight).

Reading Ranges,

  DEF, 

0 - 50% by w/w (weight),   Accuracy  +/- 0.1%

  

Calibration -

 Distilled Water

Automatic Temperature Compensation -

 68°F (20

o

C)

 

The built in temperature compensation mode is based on  

 

the temperature coefficients that have been calculated 

 

specifically for diesel exhaust fluid.

Prism -

 Glass

Illumination -

 589nm LED

Dimensions -

 54 x 27 x 100 mm / 2.13 x 1.06 x 3.9 inches

Power - 

2 AAA Batteries, included.

Battery life - 

10,000 readings, Auto Off Sleep Mode.

Ratings - 

IP65 Dustproof/Water Resistant, CE, RoHS, and WEEE  

 

  compliant.

Warranty - 

One year against manufacturing defects. Evidence of  

 

     tampering voids warranty.

ERROR CODES:

Err01 - No sample present. Add sample.

Err02 - Inadequate sample. Add additional sample.

Err03 - Sample exceeds the refractive index reading range.

Err04 - Sample temperature has not equilibrated. Allow more time for  

            temperature equilibration.

Err05 - Excessive ambient light. Cover sample well when reading.

Err06 - Excessive ambient light. Cover sample well when reading.

Err07 - Positive calibration error. Re-calibrate with distilled water.

Err08 - Negative calibration error. Re-calibrate with distilled water.

Err09 - Poor sample condition. Sample may not be able to be read.

Err10 - Sample type may not be able to be read.

Err12 - Index of sample is out of range of instrument.

Err5X - For any errors in this series contact Reichert Analytical 

            Instruments for technical assistance.

BATTERY CONDITION:

The following display examples represent either a low battery condi-

tion or an extremely low battery condition where the instrument will 

no longer operate properly. In either case the batteries should be 

promptly replaced.

 

 

      Battery condition low. 

 

 

      Replace Batteries.

 

 

      

 

 

      Battery condition extremely low. 

 

 

      Replace Batteries.