6.1 Daily functional checks
To ensure the user’s safety a daily functional check of the most vital parts should be carried
out as described below.
- Test the brakes (both operating brake and user brake),
- Check that the seat tilt and recline functions.
- Check that the wheels and cushions are correctly located and are well maintained and clean.
- Check that the antitip devices are in the appropriate position.
- If the seat is equipped with a user card, make sure that the chair’s settings are consistent with its
recommendations.
7. RECYCLING & DISPOSAL
The HD Balance can be largely recycled. Instructions for disassembly and recycling are avail-
able in the
Scrapping instructions HD balance
article number. 95735-1.
EXPLANATIONS FOR TABLE 2
COATED AND SURFACE TREATED SURFACES
AND PLASTIC PARTS
- Wash/wipe with regular all-purpose detergent
or disinfecting agents.
No caustic agents.
SCREWS AND BOLTS:
- Make sure the nuts and bolts are tight, tighten
any that are loose.
CASTOR WHEELS:
- Check that the link wheels spin freely and do
not have any visible damage. Clear away any
debris.
ANTITIP DEVICES:
- Check that the antitip devices has no visible
damage and is operating normally.
CHASSIS:
- Inspect the chassis for cracks, warping and
other defects and notify authorized personnel
if damaged.
BRAKES:
- Check that the brakes work firmly (both
operator and user brakes).
- Contact a qualified technician if adjustment
or lubrication is necessary.
TILT AND BACK RECLINE FUNCTIONS:
- Check that the controls are easy to use and
that the levers do not have any damage.
RECLINE PIN:
- Check that the recline pin is intact and sits prop-
erly. Make sure the knob is tightly screwed on.
CUSHIONS & BELTS:
- Cushions wear out and become compressed,
check them to see if any need replacing.
- Check that the belt’s Velcro and locking mecha-
nism works properly.
ARMRESTS, SIDE SUPPORTS AND CALF
SUPPORTS
- Wipe with damp cloth or with disinfectant.
TILT LOCK:
- Remove any dirt in the holes along the rod and
the spring pin where the wire is attached.
- Check that the stop rings on the tilt bar are
securely fastened.
- If the tilt bar or wire sleeve is stiff, they can be
lubricated with a few drops of oil.
- If it gives difficulty in any other way, immedi-
ately contact the authorized personnel.
TELESCOPING ALUMINUM PROFILES (e.g.
armrests) can easily be lubricated with petroleum
jelly or food grade grease.
THE LATEST VERSION of the maintenance
instruction, article number: 95 730-1 is always