Installation
Vacuum spray degassing — 06.07.2016 - Rev. B
English — 19
6.3.1
Fitting the add-on components
Install the ball valves at the device.
1. Attach the ball valve for the "WC" make-up connection.
– If you don't connect the automatic make-up, close the
"WC" connection with a G ½ " blind plug.
2. Attach the ball valve with the "ST" dirt trap at the "DC" inlet
of the degassing.
3. Attach the ball valve at the "DC" outlet of the degassing.
6.3.2
Wall mounting
Use the bores provided at the housing rear to attach the device
at the wall. Select the attachment means according to the wall
properties and the "G" weight of the device.