4 Troubleshooting

4.1 Troubleshooting

Fault

 Possible cause

 Solutions

The tip does not vibrate and no water 
comes out after powering on and 
stepping on the foot switch

Poor contact of power plug

Plug the power supply well

Poor contact of pedal switch 

Plug the foot pedal plug well.

The fuse is broken.

Replace a T1.6AL 250V fuse

Out of battery under wireless pedal mode Replace a new battery
Poor connection under wired pedal mode  Connect the wire and screw it well
 Main unit fault

Contact local distributor or the manufacturer.

The tip does not vibrate and there is 
water spray after power on and
stepping on the foot pedal.

Loose tip

Tighten the tip

The connection between the tail wire and 
the circuit board is loose.

Contact local distributor or the manufacturer.

Handpiece fault

Contact local distributor or the manufacturer.

 cable fault

Contact local distributor or the manufacturer.

The tip vibrates but there is no water 
spray after power on and stepping on the 
foot pedal.

The resistance of the diaphragm pump 
becomes larger, and the torque of the low 
water retaining capacity is insufficient

Turn up the water volume.

Impurity in the solenoid valve

Contact local distributor or the manufacturer.

Water line clogging

Contact local distributor or the manufacturer.

After power-off, there is still water spray. Impurity in the solenoid valve

Contact local distributor or the manufacturer.

Handpiece heating

Water volume is too small

Turn up the water volume.

 Device fault

Contact local distributor or the manufacturer.

Water spray is too small

Water volume is too small

Turn up the water volume

Water line clogging

Contact local distributor or the manufacturer.

Weakened tip vibration

Tip is not tightened.

Tighten the tip

Tip is loose

Tighten the tip

Broken tip Note1

Replace the tip

Outlet water temperature is too high (over 
45° C)

Thermocouple fault

Contact local distributor or the manufacturer.

The Endo file does not vibrate

The nut is not tightened

Tighten the nut

There is no air spray and water spray 
after power on and stepping on the foot 
pedal.

Loose contact of power supply plug.

Plug the power supply plug well.

Poor connection under wired pedal mode Connect the wire and screw it well
Out of battery under wireless pedal mode Replace a new battery

The nozzle does not spray gas but 
there is water spray after power on and 
stepping on the foot pedal.

Nozzle clogging

Dredge the nozzle

Handpiece clogging

Dredge the handpiece

Clogging of handpiece tail cord

Remove the tail cord from the
main unit, dredge the tail cord or replace it.

Solenoid valve fault

Contact local distributor or the
manufacturer.

There is air flow but no water spray after 
power on and stepping on the foot pedal.

The resistance of the diaphragm pump 
becomes larger, and the torque of the low 
water retaining capacity is insufficient

Turn up the water volume

Impurity in the solenoid valve

Contact local distributor or the manufacturer.

Water line clogging

Contact local distributor or the manufacturer.

Leaking air of powder tank

The O-ring on base of powder
tank is broken

Unscrew the powder tank base, take out the 
damaged O-ring, replace the O-ring of the 
same specification attached, and screw on 
the powder tank base

Broken O-ring 

Replace the O-ring

There is powder residue at the thread, so 
that the screw is not in place.

Remove the residual powder at the thread 
part.

The upper cover of powder tank is broken. Replace the upper cover of powder tank.
The thread of powder tank is broken so 
that the screw is not in place.

Replace a powder tank

Water leakage of 
handpiece

Broken O-ring of handpiece

Replace the O-ring

The air powder scaling efficiency is 
reduced.

The powder in tank is not enough.

Add powder to the tank.

Powder residue in pipe, handpiece, or 
nozzle passage

Clean the passage with a fine needle and 
blow it off with compressed air.

Touch panel pops up prompt message

No powder tank 

Check the powder tank and reinstall.

Air pressure is too high or too low

Increase the external air pressure input,
Input air pressure range requirements:
5.5bar-7.5bar(0.55Mpa-0.75MPa)

 Pick up two handpieces at a time 

Before stepping on the pedals, please pick 
up only one handpiece

 no handpiece has been picked up

Before stepping on the pedals, please pick 
up only one handpiece

The handpiece does not match the mode 
configuration

Please select the correct handpiece and 
mode

Note: if the problems cannot be solved, please contact local distributor or the manufacturer.

Summary of Contents for PT-X

Page 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Page 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Page 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Page 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Page 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Page 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Page 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Page 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Page 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Page 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Page 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Reviews: