RF emissions CISPR 11
Class B
PT X is suitable for being used in domestic establishment and in
establishment directly connected to a low voltage power supply
network which supplies buildings used for domestic purposes
Harmonic emissions IEC 61000-3-2
Complies
Voltage fluctuations/ flicker emissions IEC
61000-3-3
Complies
Guidance & Declaration - electromagnetic immunity
The model PT X is intended for use 1n the electromagnetic environment specified
below. The customer or the user of the model PT X should assure that It is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient burst IEC
61000-4-4
± 2 kV for power supply lines
± 1 kV for Input/Output lines
± 2 kV for power supply lines,
± 1 kV for interconnecting cable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
± 1 kV line to line
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5
cycle
40% Ur (60% dip in UT) for 5 cycles
70% UT (30% dip in UT ) for 25
cycles
<5% UT (>95 % dip 1n UT ) for 5
sec
<5% UT (>95% dip in UT .)
for 0.5 cycle
40% UT (60% dip in UT for 5
cycles
70% UT (30% dip in UT for 25
cycles
<5% UT (>95 % dip in UT)
for 5 sec)
Mains power quality should be that
of a typical commercial or hospital
environment If the user of the model
PT X require continued operation
during power mains interruptions, it
is recommended that the model PT X
be powered from an uninterruptible
power adapter or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels charactenstic
of a typical location in a typical
commercial or hospital environment
NOTE Ur is the ac mains voltage prior to application of the test level.
Guidance & Declaration - Electromagnetic immunity
The model PT X is intended for use in the electromagnetic environment specified below. The customer or the user of the model PT X
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
3Vrms
150kHz - 80MHz
3V/m
80MHz - 2.5GHz
3Vrms
3V/m
Portable and mobile RF communications equipment should
be used no closer to any part of the model PT X, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=[3,5/V1]×P
1/2
d=1.2×P
1/2
80 MHz - 800 MHz
d=2.3×P
1/2
800 MHz - 2.5 GHz
Where P is the maximum output power rating of the
transmitter In watts (W) according to the transmitter
manufacturer and d Is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, ª should be less than the
compliance level in each frequency range.
b
interference may occur In the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz - 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
A) Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the model PT X is used exceeds the applicable RF compliance level above, the model PT X
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the model PT X.
B) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the model PT X
The model PT X is intended for use in electromagnetic environment in which radiated RF disturbances is controlled. The customer or
the user of the model PT X can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the model PT X is recommended below, according to the maximum output
power of the
communications equipment.
Rated maximum output power
of transmitter W
Separation distance according to frequency of transmitter m
150kHz ~ 80MHz d =1.2
80MHz ~ 800MHz d =1.2
800MHz ~ 2.5GHz d =2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73