Refine PT-X Instruction Manual Download Page 3

3

Ultrosonic handpiece

HY-2L

Reusable and Durable; Please change a new handpiece when the output power 
drops significantly or does not work.

Power cord 

/

Reusable and Durable; Please replace the power cord when it is cracked or 
non-conductive

Water pipe

/

Reusable and Durable; Please change a new pipe when the water pipe is cracked 
and leaked.

Tips

refer to the packing 
list

Please change a new tip when it is worn out. For details, see point 12 of the Safety 
Precautions.

Torque wrench

/

Reusable and Durable. Please change a new one as per Instruction Manual when 
its broken.

Foot pedal

RFS05

Reusable and Durable. Please change a new one as per Instruction Manual when 
its broken.

Water bottle 

/

Reusable and Durable. Please change a new one as per Instruction Manual when 
its broken

Powder tank 

/

Reusable and Durable. Please change a new one as per Instruction Manual when 
its broken.

1.7 Device safety classification

1) Classified by operation mode: Continuous operating device
2) Type of protection against electric shock: Class 

I

3) Degree of protection against electric shock: BF type
4) Degree of protection against harmful ingress of water: Ordinary equipment (IPX0). Foot pedal is anti-drip device (IPX1)
5) Degree of safety application in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide: Equipment 
cannot be used in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.

1.8 Main technical specification

Specification 

PT X

Main Unit Size (mm)
Length*Wideth*Height

320mm330mm120mm

Main unit weight

4.0kg

Handpiece model

HY-2L / XP-1 / XP-2

Control mode 

Touch control

Water supply mode

Water supply

Power adapter input

100VAC - 240VAC   50Hz/60Hz

Input power

200VA

Main unit fuse

T1.6AL 250V

Foot pedal shell protection 
Level

IPX1

Tip output features

Frequency: 30kHz5kHz; Output main vibration offset of tip 1μm~200μm; Semi-offset force 0.1-5N; 

Output power of tip 3W-20W

Water inlet pressure

0.1bar-5bar (0.01MPa-0.5MPa)

Air inlet pressure

5.5bar-7.5bar (0.55MPa-0.75MPa)

Water temperature

≤ 45

1.9 Operation Environment

Environmental temperature: +5℃ -  +40℃ , Relative humidity: 30%  -  75%, Atmospheric pressure: 70kPa - 106kPa, Cooling water 
temperature: +5℃ -  +25℃

2 Installation 

2.1 Main Unit Diagram

Figure 1 Front view                                          Figure 2 Back view      

2.2 Touch panel

Figure 3 Touch panel diagram

Summary of Contents for PT-X

Page 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Page 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Page 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Page 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Page 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Page 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Page 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Page 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Page 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Page 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Page 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Reviews: