Output Operation Procedure

Function  

Operation

Range  

Operation

Display

Set Range

DCA 10mA

20mA%

0.000V

0.000 to 11.000V 

-00.000 to 22.000mA

DC Voltage Output

1.  Insert one end of the test leads into the V  

output jack and connect the other end to  
the input of the meter, as shown in the  
diagram.

2.  The 

symbol will appear on the 

display indicating that the calibrator is in an  
output state.

3.  Press the V/mA button to select the V function and display the V 

unit.

4.  Press the 

 buttons to select the output digits.

5.  Press the 

 button to change the numerical value of the set 

digits. Hold the button and the value will keep varying.

6.  Press the ON/OFF button turn the on and off the output followed 

by displaying the ON or OFF symbol.

DC Current Output

1.  Insert one end of the test leads into the 

+mA output jack and connect the other 
end to the input of the meter, as shown 
in the diagram.

2.  Press the V/mA button to select the mA 

function and display the mA unit.

3.  Press the mA / % button to select the mA or % function and  

display the mA or % unit, in which 0% is 4mA and 100% is 20mA.

4.  Press the 

 buttons to select the output digits.

continued ...

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for vc04

Page 1: ...ruments www com Instruction Manual Black Pantone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Pantone 123 Yellow 0 27 100 Model VC04 Voltage Current Calibrator www com information itm com 1 800 5...

Page 2: ...Description 6 7 Operating Instructions 8 11 Power ON OFF 8 Automatic Power Off 8 Output Function 8 Output Operation Procedure 9 DC Voltage Output 9 DC Current Output 9 10 25 Step Current Output 10 Cu...

Page 3: ...e following regulations Warning Do not operate the instrument in the presence of flammable or explosive gas or vapor This is extremely dangerous Never apply more than 30V between any two terminals or...

Page 4: ...year of calibration Output Range Output Range Resolution Accuracy Note DCV 10V 0 000 to 11 000V 1mV 0 05 of set value 2mV Max output current 2 mA DCA 20mA 0 000 to 22 000mA 0 001mA 0 05 of set value 4...

Page 5: ...30V between any two terminals or between any terminal and earth ground Operating temperature 0 C to 50 C Operating relative humidity 80 RH Storage temperature 10 C to 55 C Storage humidity 90 RH Safe...

Page 6: ...534 Blue 100 80 30 5 100 LOOP 30V MAX mA 25 ON OUTPUT 30V MAX 30V MAX XMT V V mA VOLTAGE mA SOURCE Instrument Description 1 LCD screen 2 Power button 3 ON OFF 4 Zero FS points select button 5 25 sing...

Page 7: ...s that the instrument is in an output state Indicates that the instrument is in a calibration state Indicates that the instrument is in a calibration state denoting that the zero point or the full sca...

Page 8: ...o automatically turn itself off after 10 minutes without any action Follow the steps below to turn the Auto Power Off function off 1 Be sure the meter is powered off 2 Press both the POWER and mA V bu...

Page 9: ...e meter as shown in the diagram 2 The symbol will appear on the display indicating that the calibrator is in an output state 3 Press the V mA button to select the V function and display the V unit 4 P...

Page 10: ...the set output to mA or and display the mA or mA unit 5 Press the button to change the output in a value of 25 in which 0 indicates 4mA and 100 indicates 20mA 6 Press the ON OFF button turn the on an...

Page 11: ...brator and connect the other end with the input terminal of the user s instrument as shown in the following diagram 2 The button operation is the same as that as indicated in the Current Output sectio...

Page 12: ...urn the power off the meter and remove any test leads that may be attached to the meter 2 Remove the holster from the calibrator by releasing the lock in the indicated direction 3 Replace the old batt...

Reviews: