Operating Instructions
Calibration Procedures
Using a standard Acoustic Calibrator
.
(94dB, 1kHz Sine wave)
1. Make the following buttons
.
and function switch settings:
Display: dB, A, Lo, F
Function: A-Weighting
Response Time: FAST
Level range: 35 to 100dB
Measurement mode: MAX Hold and
.
Data Hold Mode function disable.
2. Insert the microphone housing
.
carefully into the insertion hole
.
of the calibrator.
3. Turn on the switch of calibrator
.
(94dB @ 1kHz) and adjust the
.
CAL potentiometer of the unit.
.
The level display will indicate
.
the desired level.
This product is well calibrated when
manufactured. We recommend it be
.
recalibrated every year.