NOTE
: The meter calculates the mean of a max of 5 calibration readings.
When full, the newest calibration value will replace the oldest
calibration value. Press Blue Bar Button to exit current
calibration, or after 30 seconds the calibration will be become
effective automatically. Press the Zero Calibration Button to make
the current calibration effective. Even while a series of
measurements is being taken, foil calibration can be carried out
often as necessary. The old calibration will be overwritten;
the Zero Calibration remains in memory.
Two-point Calibration
The meter needs to be in Single Mode (see “Working Mode” for instruc-
tions) for Two-point Calibration. This method requires the use of two
different foils, a thin foil, and a thicker foil (about 1.5 times as thick as the
thin one.) For best results, the thickness to be expected should be some-
where between the two calibration values. This method is suitable for
taking measurements on rough shot-blasted surfaces or for high-precision
readings. It is recommended to take 5 measurements as to reduce the
effect of scattering, which occurs during calibration of upper and lower
values. The calibration foils may be used in any order.
1) Zero-point Calibrate according to the instructions in “Zero-point
Calibration” section.
2) One-point Calibrate according to the instructions in “One-point
Calibration” section.
3) Repeat Step 2.
4) Take readings by placing the probe on the coating to be measured
and raise it after the beep. The reading will show on the display.
NOTE
: Apply the probe to the test sample several times. The thickness of
the foil should be roughly equivalent to the estimated
coating thickness. Even while a series of measurements is being
taken, foil calibration can be carried as often as necessary. The
old calibration will be overwritten; the Zero-point Calibration
remains in memory until another is made.
www.reedinstruments.com