9
Setting the Datalogging Recording Interval
In order to set the sampling rate on the meter, follow steps 1 through 4.
1. While the meter is off, enter the Setup mode by pressing and holding
the Power Button and A/C buttons simultaneously. "Set" will flash on
the display indicating you are in the Setup mode.
2. Press the
FAsT/slOw
(INTV) button to enter Recording Interval
setup screen as indicated by "Int".
3. Press the Up and Down arrows to adjust the recording interval. A
maximum value of 1 minute and minimum value of 1 second can
be set.
4. Press the
FAsT/slOw
(INTV) button to save the desired sampling
rate and resume normal operation.
NOTE:
Press the Power button to exit Setup mode at any time.
Datalogging
1. Turn the meter on and select your desired sampling rate
(See "Setting the Datalogging Recording Interval" above).
2. Press the
REC
button to begin datalogging.
3. Press the
REC
button again to stop datalogging.
Clearing Stored Data
To delete internal memory press and hold the
REC
button and Power
button simultaneously for 5 seconds when the meter is off. The display
will show a 5 second countdown. Once the countdown is completed
"Clear" will display confirming the memory has been erased.
Calibration Procedure
continued...