7

Operating Instructions

1.  Turn the meter on by pressing the Power Button. To turn the meter 

off, press and hold the Power Button.

2.  Press the Time Weighting Button to select the desired Response 

Time. If the sound source consists of short bursts, set the  

response to FAST. To measure average sound level, select SLOW

3.  Press the Frequency Weighting Button to select the Frequency. 

Select "A" Weighting for a general noise sound level, or "C" 

Weighting for measuring high level noise. "C" Weighting is a stan-

dard weighting of the audible frequencies commonly used for the 

measurement of Peak Sound Pressure Level. If the "C" Weighted 

Level is much higher than the "A" Weighted Level then there will be a 

large amount of low-frequency noise.

4.  Press the Level Range Button to select the desired dB range. 

Choose a dB range in which the measured level is registering 

approximately in the middle of the bar graph. If the over range 

indicator appears during your measurement, the upper limit of the 

selected dB range has been exceeded. Adjust the selected dB range 

until the over range indicator disappears from the screen. Further-

more, if the under range indicator appears during your measurement, 

the lower limit of the selected dB range is too high to register your 

measurement. Adjust the selected dB range until the under range 

indicator disappears.

5.  Hold the instrument comfortably in one hand (away from your body) 

or position it on a tripod for long-term monitoring. Point the micro-

phone in the direction of the noise to take a measurement.

NOTE

: Direct contact with strong winds or blowing air on the microphone 

may cause measurement errors. These effects can be reduced by using 
the included windscreen.

MIN/MAX

Press the 

MIN/MAX

 button to display the Maximum and Minimum sound 

level values recorded during the measurement process.

continued...

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Summary of Contents for R8080

Page 1: ...Sound Level Meter Data Logger Instruction Manual R8080 REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Page 2: ...F 8 Backlight 8 Setting the Date and Time 8 Setting the Datalogging Recording Interval 9 Datalogging 9 Clearing Stored Data 9 Calibration Procedure 9 10 Battery Replacement 10 AC DC Signal Output Jack...

Page 3: ...ce will apply Safety Never attempt to repair or modify your instrument Dismantling your product other than for the purpose of replacing batteries may cause damage that will not be covered under the ma...

Page 4: ...condenser Digital Analog Bargraph Yes 30 segment Display 4 digit LCD Backlit Display Yes Min Yes Max Yes Alarm Indicators Under and Over visual on screen Datalogging Capabilities Yes Real Time Clock a...

Page 5: ...USB Cable PC Software Hard Carrying Case Batteries Instrument Description 1 Windscreen 2 Display 3 Power Backlight Button 4 MIN MAX Clock Button 5 Record Button Up 6 Frequency Weighting Setup Button...

Page 6: ...M drive from your computer Double click on the setup file and follow the on screen prompts to install the software If your computer does not have a CD drive please visit www reedinstruments com softwa...

Page 7: ...oximately in the middle of the bar graph If the over range indicator appears during your measurement the upper limit of the selected dB range has been exceeded Adjust the selected dB range until the o...

Page 8: ...ed to a PC Backlight Press the Backlight Button to turn the LCD Backlight on and off The backlight will automatically turn off after 30 seconds Setting the Date and Time In order to set the time and d...

Page 9: ...T SLOW INTV button to save the desired sampling rate and resume normal operation Note Press the Power button to exit Setup mode at any time Datalogging 1 Turn the meter on and select your desired samp...

Page 10: ...cated on screen by CAL 94dB 4 Insert the microphone into the calibrator 5 Press the Up and Down arrow buttons to increase or decrease the dB level to match the Sound Calibrator output 6 Once the desir...

Page 11: ...Connector Diagram AC Output Voltage Range 1 Vrms at FS full scale FS refers to the upper limit of the selected dB level range For example if 30 90dB is the selected level range 1Vrms 90dB DC Output V...

Page 12: ...Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or repl...

Page 13: ...ED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets p...

Page 14: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Reviews: