INSTALLATION  

 

Important- 

before you install the Rl101 h2s head make sure you have read the basic operations 

to ensure proper understanding of the product. Also check battery voltage to ensure Rl101 will function 
properly. 

When installing the Rl101 h2s Head make sure to calibrate the sensor head. Put the Head into “

cal

” 

mode. When calibration gas is applied, the unit will display the reading. Use the Up and Down buttons 
to adjust the reading to match the value of the calibration gas. ( 

# 3 in basic operations) 

 

The Rl101 Head may be mounted 18 to 24 inches above the ground with the sensor housing pointed 

down towards the ground. For best signal strength position the glass window in line of sight of the 
Redline Monitor, also make sure there are no metal/nonmetal obstructions so signal will not be blocked. 

 

 

 

 

 

MAINTENANCE 

Redline Instruments recommends you check battery voltage every 90 days and Calibration of the head 
every 30 days. Recommended battery replacement at 7.5VDC or lower, if battery is at 5.3VDC there will 
not be sufficient power to transmit a signal. If your instrument is in need of repair you may send your 
instrument to any of the redline location for repair. See website for location addresses. 

For 

Troubleshooting assistance call the Redline office at (979)776-7200.

 

 

Summary of Contents for RL-101

Page 1: ...cover tight when circuits are alive Test Instrument regularly to ensure it is operating properly IMPORTANT SAFETY WARNING Failure to follow all safety and operational instructions may result in instrument damage serious injury or death Power Consumption and Voltage rating RL101 instrument is powered by 6 AA batteries 9VDC internally 360uA continuous consumption when the device is operational INSTR...

Page 2: ...EW Modes 1 Normal Operation 2 NULL 3 Calibrate 4 Assign Head Address 5 Set Background Gas Level 6 Check Battery Voltage 7 Select Redline or WF784 Radio Message 8 Revision 9 Relay Test 10 Radio Signal Strength Input 11 Select Redline or WF784 Radio Message same as step 7 12 OFF 13 Startup ...

Page 3: ...s TEST to get to the third Mode press TEST again while in the 2nd Mode Pressing TEST while in the 3rd Mode will get back into the 2nd mode or addressing Pressing MENU multiple times will get back to Normal Operating mode If left in any mode except OFF the unit will return to Normal Mode after three minutes When the Head unit changes modes it will send a message to the Monitor Menu Reed Test Reed U...

Page 4: ...in normal mode it will display a zero 2 NULL To place the Rl101 into NULL Mode press MENU button our use the magnet with the reed switch once NUL will flash on the display The RL101 Head unit monitors the gas level but does not send messages when in NULL Mode NULL Mode is used removing any errors in the amplifiers that may cause a gas reading when there is no gas Use the Up or Down buttons to adju...

Page 5: ... Normal Operation and will begin sending messages to the monitor 4 Assign Radio Head to Channel To access to this feature press TEST once Adr will flash on the display when in that mode Use the UP and Down buttons to adjust the Address of the Head unit There must be a corresponding address set in the Monitor Press Menu button or use reed switch twice to enter CAL mode Use the Up and Down buttons t...

Page 6: ... This mode is new to revisions 2 0 or greater If the RL101 sensor head is in the wrong setting of 784 use the up or down buttons to change back to Red if the RL101 sensor head is in 784 it will not communicate to the Redline Monitor 8 Revision This mode shows the revision of the software that is being used in the RL101 Head unit To access press TEST Twice 9 Relay Test To place the RL101 sensor Hea...

Page 7: ...erformed for at least one second The Head unit can be moved away from obstructions up or down or pointed to get the best reading When the MENU button is pressed to leave this mode the last reading taken will be stored in memory the stored reading will be included in the messages to the Monitor so this reading can be examined at the Monitor If no Monitor is in range the readings will be very low or...

Page 8: ...to 24 inches above the ground with the sensor housing pointed down towards the ground For best signal strength position the glass window in line of sight of the Redline Monitor also make sure there are no metal nonmetal obstructions so signal will not be blocked MAINTENANCE Redline Instruments recommends you check battery voltage every 90 days and Calibration of the head every 30 days Recommended ...

Reviews: