10. Radio Signal Strength Input 

To get to this mode press the TEST button twice and the MENU button twice "

rIn

" will then flash 

in the display. This mode allows the Head unit and Monitor to exchange messages (this mode is disabled 
when in the 784 mode since the 784 does not resend any messages). When the Head unit sends a  
message, the Monitor knows that the Head unit is in this mode and resends the message. The  
Head unit stays awake in order to receive this resent message. The Head unit measures the  
strength of the signal and displays it (the maximum Value is 255); the units should work  
properly until the signal level drops to about 50. 

Caution

: since the radio stays awake for this test, the battery is being discharged much faster in this 

mode, so try to spend as little time in this test as possible.) 

 The Head unit transmits a message once per second and the Monitor responds immediately, so 

the test must be performed for at least one  
second. The Head unit can be moved away from obstructions, up or down or pointed to get  
the best reading. When the MENU button is pressed to leave this mode, the last reading  
taken will be stored in memory; the stored reading will be included in the messages to the  
Monitor, so this reading can be examined at the Monitor. If no Monitor is in range, the  
readings will be very low or zero.  

11. OFF 

To put the Head unit in the

 OFF

 mode, press the TEST button for about 4 seconds. The  

display will show "

OFF

". 

 This mode disables all transmissions by the Head unit. The unit will transmit a message every 5  
minutes with no gas present or every 5 seconds when the gas level is equal to or greater  
than the background setting. To reduce the battery discharge, the unit can be set in 

OFF

  

mode, where it will never transmit. To wake the Head unit, press MENU, Up or TEST, and the unit will 
begin the start up sequence. 
 

12. Startup 

When the RL101 is powered or awakened, it will display several values, like gas type,  

address

 and 

battery voltage

, then the unit will count up to 30 seconds and go into 

normal operation

 

 

 

 

 

 

 

 

Summary of Contents for RL-101

Page 1: ...cover tight when circuits are alive Test Instrument regularly to ensure it is operating properly IMPORTANT SAFETY WARNING Failure to follow all safety and operational instructions may result in instrument damage serious injury or death Power Consumption and Voltage rating RL101 instrument is powered by 6 AA batteries 9VDC internally 360uA continuous consumption when the device is operational INSTR...

Page 2: ...EW Modes 1 Normal Operation 2 NULL 3 Calibrate 4 Assign Head Address 5 Set Background Gas Level 6 Check Battery Voltage 7 Select Redline or WF784 Radio Message 8 Revision 9 Relay Test 10 Radio Signal Strength Input 11 Select Redline or WF784 Radio Message same as step 7 12 OFF 13 Startup ...

Page 3: ...s TEST to get to the third Mode press TEST again while in the 2nd Mode Pressing TEST while in the 3rd Mode will get back into the 2nd mode or addressing Pressing MENU multiple times will get back to Normal Operating mode If left in any mode except OFF the unit will return to Normal Mode after three minutes When the Head unit changes modes it will send a message to the Monitor Menu Reed Test Reed U...

Page 4: ...in normal mode it will display a zero 2 NULL To place the Rl101 into NULL Mode press MENU button our use the magnet with the reed switch once NUL will flash on the display The RL101 Head unit monitors the gas level but does not send messages when in NULL Mode NULL Mode is used removing any errors in the amplifiers that may cause a gas reading when there is no gas Use the Up or Down buttons to adju...

Page 5: ... Normal Operation and will begin sending messages to the monitor 4 Assign Radio Head to Channel To access to this feature press TEST once Adr will flash on the display when in that mode Use the UP and Down buttons to adjust the Address of the Head unit There must be a corresponding address set in the Monitor Press Menu button or use reed switch twice to enter CAL mode Use the Up and Down buttons t...

Page 6: ... This mode is new to revisions 2 0 or greater If the RL101 sensor head is in the wrong setting of 784 use the up or down buttons to change back to Red if the RL101 sensor head is in 784 it will not communicate to the Redline Monitor 8 Revision This mode shows the revision of the software that is being used in the RL101 Head unit To access press TEST Twice 9 Relay Test To place the RL101 sensor Hea...

Page 7: ...erformed for at least one second The Head unit can be moved away from obstructions up or down or pointed to get the best reading When the MENU button is pressed to leave this mode the last reading taken will be stored in memory the stored reading will be included in the messages to the Monitor so this reading can be examined at the Monitor If no Monitor is in range the readings will be very low or...

Page 8: ...to 24 inches above the ground with the sensor housing pointed down towards the ground For best signal strength position the glass window in line of sight of the Redline Monitor also make sure there are no metal nonmetal obstructions so signal will not be blocked MAINTENANCE Redline Instruments recommends you check battery voltage every 90 days and Calibration of the head every 30 days Recommended ...

Reviews: