Publication No. REF-000-009-R00
Red Rapids
Page 8
3.2 Performance
Performance may vary depending on the quality of the power
supply and EMI environment of the host.
3.2.1 AC-Coupled Performance
Parameter
Min
Typ
Max
Unit
Passband
(1)
1 dB
3 dB
1
150
0.1
400
MHz
MHz
SNR
20.17 MHz Input
70.17 MHz Input
(2)
125.17 MHz Input
(3)
74.9
74.1
71.1
dBFS
dBFS
dBFS
SINAD
20.17 MHz Input
70.17 MHz Input
(2)
125.17 MHz Input
(3)
74.8
74.1
71.0
dBFS
dBFS
dBFS
SFDR
20.17 MHz Input
70.17 MHz Input
(2)
125.17 MHz Input
(3)
95
89
87
dBc
dBc
dBc
Channel to Channel Isolation
25 MHz
50 MHz
75 MHz
100 MHz
>90
85
84
81
dB
dB
dB
dB
Notes:
(1)
Measured across band using ADC output.
(2)
Performance extrapolated from -4.8 dBFS plot due to test equipment limitations.
(3)
Performance extrapolated from -2.0 dBFS plot due to test equipment limitations.