background image

4

Note: The nominal output value for the various output ranges are designated in the 

following order:  (0 to 20 mA, 4 to 20 mA, 0 to 10 VDC)

This calibration procedure can be used to assign the high and low input values if a 

thermocouple calibrator is not available.

Note: To abort this calibration and reset to the previous settings, set the FIELD CAL 

switch(#2) OFF prior to the final OFF setting of the OUTPUT CAL switch (Step 

2.12) and turn off power. Wait 5 seconds and then turn on power and the previous 

settings will be loaded.

2.1  Connect the accurate Adjustable Millivolt Source to the TC input terminals.

2.2  Set the ICE PT EN/DIS switch (#4) ON to disable Ice Point Compensation.

2.3  Set the Type and Range for the thermocouple or mV used in your application 

(DIP switches #6 through #10).(TC type “J”, Range 0 shown)

2.4  Set the FIELD CAL switch (#2) ON.[Output goes to -0.8 mA, 3.5 mA, or -0.4 V 

nominal]

2.5  Apply the input signal (mV equivalent for the thermocouple temperature) for the 

analog output low value.

2.6  Set the OUTPUT CAL switch (#1) ON. [Output stays at -0.8 mA, 3.5 mA, or -0.4 

V nominal]

2.7  Adjust the input signal up until the output equals desired low value.
2.8  Set the OUTPUT CAL switch (#1) OFF. [Output increases to 22.9 mA, 22.5 mA, 

or 11.5 V nominal]

2.9  Apply the input signal (millivolt equivalent for the thermocouple temperature) for 

the analog output high value.

2.10 Set the OUTPUT CAL switch (#1) ON. [Output decreases to 21.1 mA, 20.7 mA, 

or 10.6 V nominal]

2.11 Adjust the input signal down until the output equals desired high value.
2.12 Set the OUTPUT CAL switch (#1) OFF.

2.13 Set the FIELD CAL switch (#2) OFF.

2.14 Set the ICE PT EN/DIS switch (#4) OFF to enable Ice Point Compensation.

2.15 Disconnect millivolt source from the ITMA and connect the actual sensor to be 

used in the application.

ERROR

POWER

VDC+

10

TC+

7

V+

4

I+

1

VDC-

TC-

12

11

9

8

6

5

V-

3

2

I-

POWER SUPPLY

9 to 32 VDC

+

-

VOLTAGE METER

OR

CURRENT METER

ADJUSTABLE

MILLIVOLT

SOURCE

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

ON

1 2 3 4 5 6 7 8 9 10

4

1 2 3

ON

6

5

7 8 9 10

ON

1 2 3 4 5 6 7 8 9 10

Field Calibration Wiring

Step 2.2 & 2.3

Step 2.4

Step 2.6

Step 2.8

Step 2.10

Step 2.12

Step 2.13

Step 2.14

2.0 Field Calibration With an Accurate Adjustable Millivolt Source: (Alternate Method)

Note: The unit needs to have the Field Calibration completed by the operator before 

normal operation. To abort this calibration and reset to the previous settings, set 

the FIELD CAL switch(#2) OFF prior to the final OFF setting of the OUTPUT 

CAL switch (Step 1.13) and turn off power. Wait 5 seconds and then turn on power 

and the previous settings will be loaded.

Field Calibration with a Thermocouple Calibrator

1.1  Enable the Ice Point Compensation on the Thermocouple Calibrator and set it to 

the Thermocouple type being used in your application.

1.2  Connect the thermocouple wire as selected in step 1 to the TC input terminals of 

the ITMA and the thermocouple calibrator.

1.3  Set the ICE PT EN/DIS switch (#4) OFF to enable Ice Point Compensation.

1.4  Set the Type and Range for the thermocouple or mV used in your application 

(DIP switches #6 through #10). (TC type “J”, Range 0 shown)

1.5  Set the FIELD CAL switch (#2) ON. [Output goes to -0.8 mA, 3.5 mA, or -0.4 V 

nominal]

1.6  Apply the input signal for the analog output low value.

1.7  Set the OUTPUT CAL switch (#1) ON. [Output stays at -0.8 mA, 3.5 mA, or -0.4 

V nominal]

1.8  Adjust the input signal up until the output equals desired low value.

1.9  Set the OUTPUT CAL switch (#1) OFF. [Output increases to 22.9 mA, 22.5 mA, 

or 11.5 V nominal]

1.10 Apply the input signal for the analog output high value.

1.11 Set the OUTPUT CAL switch (#1) ON. [Output decreases to 21.1 mA, 20.7 mA, 

or 10.6 V nominal]

1.12 Adjust the input signal down until the output equals desired high value.

1.13 Set the OUTPUT CAL switch (#1) OFF.

1.14 Set the FIELD CAL switch (#2) OFF.

1.15 Disconnect the thermocouple calibrator from the ITMA and connect the actual 

sensor to be used in the application.

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

1

6

5

4

3

2

10

9

8

7

ON

Step 1.13

Step 1.3 & 1.4

Step 1.5

Step 1.7

Step 1.9

Step 1.11

Step 1.14

OUTPUT CAL

FIELD CAL

BASIC CAL

ICE PT DIS/EN

OPEN SEN DN

RANGE

RANGE

DIP SWITCH SETTINGS:

ON = 1

OFF = 0

TYPE

TC
TC TYPE

MODEL ITMA

9-32 VDC

ERROR

POWER

VDC+

10

TC+

7

V+

4

I+

1

VDC-

TC-

12

11

9

8

6

5

V-

3

2

I-

4

2

1

3

7

5

6

8

10

9

ERROR

POWER

VDC+

10

TC+

7

V+

4

I+

1

VDC-

TC-

12

11

9

8

6

5

V-

3

2

I-

POWER SUPPLY

9 to 32 VDC

+

-

VOLTAGE METER

OR

CURRENT METER

THERMOCOUPLE

CALIBRATOR

Field Calibration Wiring

Note: The nominal output value for the various output ranges are designated in the following order:  (0 to 20 mA, 4 

to 20 mA, 0 to 10 VDC)

Allow a 30 minute warm-up period before starting Field Calibration. Field Calibration scales the voltage or current 

output to a temperature or mV input. This procedure assigns an input value to analog output low and an input value 

to analog output high. The microprocessor handles configuring the output so it is linear to the temperature or mV input. 

The Field Calibration procedure is described below.

1.0 Field Calibration

CALIBRATION PROCEDURES

OUTPUT CAL

FIELD CAL

BASIC CAL

ICE PT DIS/EN

OPEN SEN DN

RANGE

RANGE

DIP SWITCH SETTINGS:

ON = 1

OFF = 0

TYPE

TC
TC TYPE

MODEL ITMA

9-32 VDC

ERROR

POWER

VDC+

10

TC+

7

V+

4

I+

1

VDC-

TC-

12

11

9

8

6

5

V-

3

2

I-

4

2
1

3

7

5

6

8

10

9

Summary of Contents for ITMA

Page 1: ...ment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired SPECIFICATIONS 1 POWER 9 to 32 VDC 1 75 W The power supply must have 300 mA for 200...

Page 2: ...rols and heaters etc The cables should be run through metal conduit that is properly grounded This is especially useful in applications where cable runs are long and portable two way radios are used i...

Page 3: ...e case and cut the wire jumper The jumper is located to the left side of the board as shown in the drawing POWER Connect DC power to terminals 10 and 12 observing proper polarity Be certain DC power i...

Page 4: ...en turn on power and the previous settings will be loaded Field Calibration with a Thermocouple Calibrator 1 1 Enable the Ice Point Compensation on the Thermocouple Calibrator and set it to the Thermo...

Page 5: ...Output increases to 0 8 mA 3 5 mA or 0 4 V nominal or more 4 19 Perform a Field Calibration See Section 1 0 1 2 3 4 5 6 7 9 8 10 ON 1 2 3 4 5 6 7 9 8 10 ON 1 2 3 4 5 6 7 9 8 10 ON 1 2 3 4 5 6 7 9 8 1...

Page 6: ...S POWER SUPPLY 24VDC COMM MASTER CONTROL ROOM G Rail Installation To install the ITMA on a G style DIN rail angle the module so that the upper groove of the foot catches under the lip of the top rail...

Page 7: ...7 This page intentionally left blank...

Page 8: ...Buyer its employees or sub contractors are or may be to any extent liable including without limitation penalties imposed by the Consumer Product Safety Act P L 92 573 and liability imposed upon any p...

Reviews: