RF emissions
CISPR 11
Class B
The
Model 760R
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Model 760R - Electromagnetic Immunity – Manufacturer’s Declaration
The
Model 760R
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Model 760R
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
Complies
Mains power quality should be that of a
typical domestic or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
Complies
Mains power quality should be that of a
typical domestic or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
>95% dip in UT for
0.5 cycle
60% dip in UT for
5 cycles
30% dip in UT for
25 cycles
>95% dip in UT for
5 cycles
Complies
Mains power quality should be that of a
typical domestic or hospital
environment.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical domestic l or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Model 760R - Electromagnetic Immunity - Manufacturer's Declaration
The
Model 760R
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Model 760R
should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
Model
760R
, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
L98107R0-G
37
Summary of Contents for RP Lite Model 760R
Page 2: ......
Page 4: ......
Page 12: ...RP Lite Model 760R RP Sports User Manual Figure 1 The Massage Sequence L98107R0 G 8...
Page 17: ...L98107R0 G 13...
Page 20: ...RP Lite Model 760R RP Sports User Manual Step 1 Set up the RP Lite Console L98107R0 G 16...
Page 27: ...Step 4 Connect the Garment Hose Bundle to the Console L98107R0 G 23...
Page 34: ...RP Lite Model 760R RP Sports User Manual Replacing the Battery L98107R0 G 30...
Page 36: ...RP Lite Model 760R RP Sports User Manual 9 Troubleshooting L98107R0 G 32...