
Operating instructions
Peristaltic Pump APY Series
17
7.4.1 Discharge side
- The discharge line is to be kept as straight and short as
possible, in order to avoid performance reduction.
- The diameter must correspond to the rated diameter of the
tube. A larger diameter is recommended in the event of viscous
liquids.
- Do not close or block the outlet since, as it is a volumetric
pump, it would cause the tube to break or an electro-
mechanical overload.
8
TORqUE VALUES
8.1 Lubrication
APY pumps only require a thin film of grease to reduce the friction
between the roller and the outer surface of the tube. No other lubrication
is necessary.
Lubricant quantity table
DESCRIPTION
UNITS
APY
A
Cover bolts
–
Manual,
no bolts
B
UNIT
APY
Lubricant
–
realax
Silicone grease
quantity
gramme
10-20
ounce
0.35 – 0.71
A
B
Summary of Contents for APY
Page 1: ...Operating instructions APY Series Peristaltic pump ALL THE HOSE PUMP YOU LL EVER NEED...
Page 25: ...25 25 Troubleshooting Spare parts End of life...
Page 30: ...Operating instructions Peristaltic Pump APY Series 30 10 2 Spare parts exploded APY LAB Model...
Page 33: ...33 12 Declaration of Conformity K t d d W Z Wz W d Wz Z W d D d W W W...
Page 39: ...Operating instructions Peristaltic Pump APY Series 39...
Page 40: ...AHQ IOM EN 0971 0521 www realaxpumps com ALL THE HOSE PUMP YOU LL EVER NEED...