RBR#0008819revB
8
1.
2.
3.
1.
2.
3.
Disconnecting the umbilical cable
Locate the small latch on the side of the connector.
Press the latch with your thumb and pull the connector down.
Set the battery end-cap aside..
.
.
.
Closing the instrument
If disconnected, insert the umbilical cable back in its socket.
Align the battery end-cap with the slot on the instrument housing and gently push down to ensure it fits in place.
Reinstall the two cap screws and tighten to 1/4 turn past tight (10Nm max).
USB-C port
Do not pull on the wires attached to the connector!
3.2 RBR
quartz³
BPR
|
zero interface
The RBR
quartz³
BPR
|
zero instrument provides an internal USB-C port and several external communication options.
Select from RS-232, RS-485, and Ethernet at the time of order, and RBR will wire the battery end-cap to support your
preferred external connection.
.
Patch cables and underwater extension cables are sold separately.
.
USB-C connection
The USB-C port is located under the battery end-cap.