background image

RBR#0008819revA

 –  

3

2  Introduction

You can access the RBR instrument guide on the USB data stick provided when you purchase a logger, from the Help 

menu in Ruskin, and on the RBR web site, at 

www.rbr-global.com

.

Summary of Contents for quartz3 BPR zero

Page 1: ...rbr global com INSTRUMENT GUIDE BPR zero...

Page 2: ...9 6 2 Changing the batteries RBRquartz BPR zero 11 6 2 1 Steps 11 6 3 Installing desiccant RBRquartz BPR zero 12 6 4 Maintenance and repairs 12 6 4 1 O rings RBRquartz BPR zero 12 Inspecting O rings 1...

Page 3: ...0008819revA 3 2 Introduction You can access the RBR instrument guide on the USB data stick provided when you purchase a logger from the Help menu in Ruskin and on the RBR web site at www rbr global c...

Page 4: ...RBR 0008819revA 4 3 Revision history Revision No Release Date Notes A 04 May 2020 Original...

Page 5: ...by us The liability of RBR Ltd extends only to the replacement cost of the instrument The customer will bear all costs of shipment to us for repair all other costs including return shipment will be b...

Page 6: ...rements are critical The RBRquartz BPR zero may be supplied with external power from RBRfermata power or RBRcervata power and memory canisters or cabled to an observatory for external power and realti...

Page 7: ...35 C The instrument has 8 X AA batteries installed however this is intended only for purposes of sustaining the instrument during external power interruptions of a few minutes While the Paros will con...

Page 8: ...2000 4000 7000 dbar Initial accuracy 0 01 FS full scale Resolution 10ppb at 16Hz sampling rate Accelerometer optional Range 3g Resolution 100ng The full mechanical drawing is available in Appendix A...

Page 9: ...shown below Only the battery end cap needs to be opened to access the data connector batteries desiccant and waste reservoir To open the logger battery end cap the two M6 screws will need to be remov...

Page 10: ...il the battery end cap is mated fully to the housing Remember to keep the O rings clean and avoid scratching the O ring mating surfaces Carefully inspect the O rings as described in the Inspecting the...

Page 11: ...ent during external power interruptions of a few minutes While the Paros will continue to run and be measured the valve will not operate 6 2 1 Steps Remove the battery end cap from the logger see Open...

Page 12: ...desiccant capsules can be opened the desiccant removed and replaced or dried 6 4 Maintenance and repairs 6 4 1 O rings RBRquartz BPR zero The single most important item of maintenance on any RBR subme...

Page 13: ...bishment please contact us for advice Replacing an O ring Lever the O ring from its groove Use a soft plastic or wood tool do not use a metal screwdriver or any other tool which may scratch the surfac...

Page 14: ...easure the pressure inside the housing of the instrument which is monitored by a barometer in order to correct for the drift of the sensors over time This actuation of this valve results in a small vo...

Page 15: ...towel around the reservoir in a clockwise manner when viewed as below The towel should only wrap around the reservoir once any additional layers will make it difficult to insert 8 While holding the to...

Page 16: ...closely as possible to the current market value based on the original purchase price less depreciation Please contact RBR if you have any questions about the instrument value Send the unit using your...

Page 17: ...RBR 0008819revA 17 7 Appendix 7 1 Appendix A RBRquartzBPR zero Product Drawing...

Page 18: ...RBR 0008819revA 18...

Page 19: ...RBR 0008819revA 19 7 2 Appendix B RBRquartzBPR zero Service Drawing...

Page 20: ...RBR 0008819revA 20 7 3 Appendix C RBRquartzBPR zero Connector Details...

Reviews: