5. Failure Display  

Instrument  failures  will  be  displayed  as  bars  in  the  LCD  display.  By  confirming  with  the  “I/O” 
button the instruments may be operated nevertheless. The failure will be displayed in exchange 
to the normal display. (bar 14 is the very right one) 
 
 

Failure Discription

Failure Discription

Failure Discription

Failure Discription    

      Failure reason

      Failure reason

      Failure reason

      Failure reason    

Bar 14 (wrong calibration) 

• 

when making the zero calibration a gas 

concentration was present 

• 

to less difference between zero and 

1000ppm calibration 

• 

Sensor defect 

• 

 

Bar 11 (Sensor head) 

    

    

    

    

• 

no sensor head present 

• 

Heating element of the sensor defect 

    

Bar 8 (wrong sensor head) 

 

 

 

    

• 

Wrong sensor head has been mounted  

(LSK head at LSG instrument)    

• 

 

 see setting of the soldering bridges    

Bar 5 (EEPROM data failure) 

    

    

    

    

• 

EEPROM contains no valid calibration data 

• 

 failure appears when first switch on after 

mounting until calibration  

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for ecom LSG

Page 1: ...von Carsten Pauli E Mail pauli rbr de Erstellungsdatum 22 02 2011 rbr Messtechnik GmbH Am Gro en Teich 2 D 58640 Iserlohn Tel 49 0 2371 945 5 Fax 49 0 2371 40305 Internet www rbr de E Mail info rbr de...

Page 2: ...phase 4 2 2 Measurement 4 2 3 Scale of the bar graph display 5 2 4 Sensitivity 5 2 5 Automatic Switch Off 5 2 6 Battery 5 2 7 Operation via the Jack Plug 6 3 Calibration 6 3 1 Change in the software f...

Page 3: ...O button the instrument is switched on and the display mode can be changed By long pressing of the I O button the LSG LSK will be switched off Press the button as long as the bars cannot be seen anym...

Page 4: ...the consumption increases to 180mW 20mW Attention The energy consumption is pulse shaped 1 7 Beep The beep will sound for function control during switch on The ppm beep can be switched off optionally...

Page 5: ...when the instrument is switched off It is done when the instrument is switched on a warm up phase will be run automatically During the first 10 seconds of the warm up phase the voltage consumption of...

Page 6: ...ng and factor 5 right bars are blinking At factor 5 ten bars will be shown already at a concentration of 200 ppm 2 5 Automatic Switch Off Should no button be pressed the instrument will switch off aft...

Page 7: ...zero calibration The detectable gas may not be in the ambient air The zero calibration will be done automatically after every warm up phase The resistance of the sensor must be in the range of 0 333 t...

Page 8: ...the humidity probe and both right connections are for the instruments which should be calibrated humidity probe connect gas bottle LSG with 1000ppm Methane in synth air LSK with R134A in synth air ad...

Page 9: ...son Failure reason Failure reason Failure reason Bar 14 wrong calibration when making the zero calibration a gas concentration was present to less difference between zero and 1000ppm calibration Senso...

Page 10: ...is switched Instrument Instrument Instrument Instrument Sensor Sensor Sensor Sensor app sensor resistance app sensor resistance app sensor resistance app sensor resistance in air in air in air in air...

Page 11: ...7 Testing Bottle A mixture is made of 40 parts water and one part Ethanol Ethanol 652 EG No 200 578 8 Of this mixture 8 ml will be filled in the cotton of the bottle The closing of the bottle has to...

Page 12: ...tly to the Cyclone Pro and contains all settings When programming t When programming t When programming t When programming the sensor head may not be populated he sensor head may not be populated he s...

Page 13: ...button menu than press the start button on the Cyclone Pro transfer starts 9 Exchange the Sensor unit The customer and all service department could exchange the sensor unit The sensor unit is pre cali...

Page 14: ...ion Even after the warm up phase is finished a new zero point is determined cyclical and the linearization is calculated new when having a positive sensor drift resistor bigger than to the end of the...

Reviews: