Compass
Standard 22
Operator Manual
Compass Standard 22
and
Standard 22 Compact
41
4102.DOC010102
Edition: February 2015
-- Open the ventilation outlet (Figure 15
/1)
, the supporting liquid screw (Figure 15
/3)
and the distilled water screw ( Figure 15
/2)
.
-- Fill in distilled water (
only
distilled water from Raytheon Anschütz,
part number 148--398, one 230cm
3
bottle) into the inlet for
supporting liquid until the measuring cone shows “Level O.K.”, see Figure 16 .
-- Fill in the remaining content of the bottle into the inlet for distilled water until the
Dip--Stick shows a level according to Figure 17 .
-- Close all in-- and outlets, if applicable use new sealing rings.
-- Place the outer sphere in the compass enclosure.
--
Pay attention to the two locking pins on the
closure.
-- Hold the outer sphere with both hands and guide it under the snap closures on the
pendulum joint.
-- Using the thumbs, press down and engage the two snap closures.
-- Carefully turn the outer sphere through 90
and repeat the process with the other two
snap closures. The outer sphere is now fastened to the pendulum joint.
-- Switch on the follow--up function again (DIP switch 7 into the lower position,
see Figure 18 ).
-- Close the door of the enclosure and fasten it with screws, paying attention to the
earthing strip.
After this procedure is finished, it is strongly recommended to
observe the heading for a while (30 minutes), because of
stimulation of the gyrosphere (caused by a slight,
unintentional horizontal inclination).