5 Quick Start Guide
5.1 Operation
This section describes a typical, step by step, operation of the Micro Junior 2
Open the top protection lid of the case and attach the test leads to the test set.
The Two (2) connections on the front plate marked
+I
& -I are the current leads.
The Two (2) connections on the front plate marked
+P
& -P are the Potential or Voltage leads.
By connecting a test specimen to the Current leads and applying the potential leads ( See figure )
across the specimen the resistance readings will be determined.
Turn the Power switch to the “ON” position.
The Instrument will make the initial
calibration check. Then the Initial Range
selection screen will appear.
Select the current or resistance range
expected to be measured with the rotary
knob.
(Selecting the cross current, X10A & X1A measurement range.)
Press the rotary knob when the indicator is pointing at the resistance and current level to be
measured (If unsure of current level, then start at the 1 mA range and increase the current range
until a valid reading appears).
After the current range is selected press the rotary knob again. The following measuring modes
can then be accessed:
Single
,
Cont...
(Continuous),
Auto
(Automatic),
Inter...
/Interval).
90081-2.11 Micro Junior 2
Page 10 of 44
WARNING !
WARNING !
Before Operating this or any other test equipment, read all safety warnings
and understand them fully