16
RaySafe i2 Installation and maintenance manual – Installation
Check the box Show accumulated procedure dose to show a column with the accumulated dose since last tap
on the reset button in the lower right corner of the online view. the accumulated dose per examination will also
be reset when restarting the real time display.
NOTE
The accumulated dose per examination is only recorded in the real time display memory. The
values recorded in the dosimeter is not affected.
Check the box Show units in rem instead of Sv if you want to view all measurements in rem. (Conversion factor
1 Sv = 100 rem is used.)
doSImeteR CommUnICAtIon RAnge menU
Figure 10. The dosimeter communication range menu.
In lab is the default communication range setting. If you have problem with the communication between
dosimeter and real time display, change the range.
• If you have several real time displays and dosimeters show up on an undesired display, select nearby.
• If you have trouble with dosimeters
not
showing up on the desired display, try the maximum Range setting.