25
RaySafe i2 Installation and maintenance manual – System function tests
Figure 15. The real time display Online View
CommUnICAtIon RAnge teSt
Check that the dosimeter communication range configuration is correctly tuned, that is, dosimeters stored
away from the real time displays shall not appear on the real time displays online view, but will appear when
they approach the display. See section “dosimeter communication range menu” on page 16.
NOTE:
Perform next steps only when multiple real time displays are installed.
1. visual check that dosimeter is shown on real time display ‘A’.
2. move dosimeter within range of real time display ‘b’.
3. visual check that dosimeter remains visible on real time display ‘A’ for three minutes after the move.
4. visual check that dosimeter is shown on real time display ‘b’ within 30 seconds after it has reached within
range.