User’s guide Ytterbium Pulsed Fiber Laser RFL-P200 Ver.: 1.0
24
Table 10 RS-232 Serial Port Definitions
PIN NO.
DESCRIPTION
2
RX
3
TX
5
GND
The other pins are of no use.
RS-232 serial port is used for Internal debugging.
4.4.4
Operation Regulations
4.4.5.1 Pre-inspection
1)
Before you start the laser, ensure that the supply and return connections are correct, and
confirm that there is no leakage in all the water circuits. Any abnormal condition in the water
circuits may cause a failure to the operation of the laser(See section 3.2).
2)
Make sure the device appearance is in good condition and the output fiber is neither
excessively bended nor broken.
3)
Please inspect the output head carefully for dust or other contaminations. Use appropriate
lens paper to clean it if necessary. Do not touch the output lens at any time.
4)
Make sure signal line of laser and cleaning system are properly connected.
5)
Before you start the laser, Ensure that the emergency stop button is in release
.
Turn-on procedures
Connect the laser with the AC power supply, switch on the circuit breaker, then turn the key
switch clockwise to the “ON” position. The Power Indicator will illumine, indicates the control
system is started. Wait a few seconds and press down the ”LASER” button. The illuming of the
indicator on the “LASER” button reveals the power module in the laser is on.
2)
Laser checking by cleaning application