Wuhan Raycus Fiber Laser Technologies Co., Ltd.
User Guide of RFL-C100~RFL-C2000S
41
Table 14
Instructions and solutions for alarms of laser
Alarm name
Alarm instructions and solutions
System Timer Alarm
Instruction:
The internal clock of laser is abnormal.
Solution:
If it displays this alarm, please contact Raycus.
Laser On Button Alarm
Instruction:
This alarm occurs when the “LASER” button on the front panel of the
laser has been pressed before the laser was powered on.
Solution:
Please power off the laser and pop up the ”LASER” button. Re-power
the laser and check if the alarm is released. If this alarm continues to
occur, please contact Raycus.
InterLock Alarm
Instruction:
Occurs when the InterLock is disconnected.
Solution:
Short the InterLock pins and restart the laser to try. If this error
continues to occur, please contact Raycus.
Current Driver Alarm
Instruction:
Current Driver Alarm, this error occurs when the constant current
driver board inside the laser is abnormal.
Solution:
Restart the laser, If this error continues to occur, please contact Raycus
Laser Out Alarm
Instruction:
1. Laser Out Alarm immediately occurs when the laser is powered.
Solution:
The laser will autolock after two times of this error occurs successive,
please contact Raycus
2. The laser will get this error when it does not emit with correct
settings.
Solution:
Close the shutter and pop up the “LASER” button on the front panel of
the laser to check if the red guide light is output normally when it isn’t
emitting. If there is no red guide light, please stop using the laser
immediately and contact Raycus. If the red guide light output
normally, please restart the laser to try, if the alarm continues to occur,
please contact Raycus.
Laser Power Alarm
Instruction: