Figure 14 The process of entering a control mode.
The control modes can be divided into AD mode, serial port communication
mode and RS232 mode (AD+ command) according to the different control modes
of the laser output and power. In the serial port communication mode, you can
choose to send commands or external AD analog voltage Settings for power
control.
The laser is AD mode when PIN 7 and 9 of the control interface are connected
to 24V (high level); When PIN7 and 9 of the control interface are suspended or
connected to 0V (low level), the laser enters the serial port communication mode.
When the key switch is placed in REM, the laser enters the external control mode.
At this time, the powering on and off of the laser is controlled by the modulated
signal of the control interface. When the key switch is ON, it enters the internal
control mode. At this time, the laser’s power on and off is controlled through the
serial port communication command.
4.7
Serial Port Communication Mode
In the serial port communication mode, according to communication protocol
provided by RAYCUS, users can type in instruction through PC. Then operations
such as setting parameters, reading laser state, and laser emission control can be
executed. RAYCUS provides a software that we developed for users to check the
laser state and debug simply. The installation environment of the laser client
control software should be within the WINDOWS system and Microsoft. NET
Framework 4.0.