Wuhan Raycus Fiber Laser Technologies Co., Ltd
User Guide RFL-A2000D-3000D Ver.1.0
22
MODULATI
ON
36
MOD+
Red
The cable is for modulation signal input, 24V level.
\
37
MOD-
Black
\
PE
\
Protective earth, connected to the shielding net of the
cable.
INTERLOCK
38
ITL_A
Red
Interlock
Make sure
that the
interlock is
shorted
before turn
on the laser.
39
ITL_B
Black
RS-232
40
RX
Red
Connect to the RX of upper computer.
The cable is
for RS-232
and Hyper
terminal
control.
41
TX
Yellow Connect to the TX of upper computer.
42
GND
Blue
Ground of the RS-232 interface.
\
PE
\
Protective earth, connected to the shielding net of the
cable.
CAUTION
:
Please check the control voltage level and ensure that the
level is in accordance with the requirements. Over voltage and voltage
ripple may damage the product.
4.5 Control Modes
There are two different modes of laser emission: CW and pulsed. In the CW mode, the laser
emits continuously, and you can set the output power by percentage of output power. In the pulse
mode, the laser emits in a waveform of pulses.
We provide three modes to control the product: HyperTerminal, RS-232 and AD. The hyper
terminal mode is a local control mode; RS-232 and AD mode are remote control modes. In the
HyperTerminal mode and RS-232 mode, you can set all the parameters in the CW or pulse mode,
but in the AD mode, the output power is controlled by analog signal, and the emission is
controlled by remote modulation signal.