Intended Use 

 
This water bath is intended for the purpose of heating water only. The operator may load a suitable article onto the mesh platform for the purpose of 
controlled temperature application. The water bath is intended for use with tap water or filtered water only.  
 
The water bath is 

not

 intended for use with any other medium. The water bath is 

not

 intended to provide direct-contact between the medium 

(water) and an unprotected medical, food, biological or medico- techno item. In all cases the item should be enclosed in a suitable vessel as described 
below. 
 

Suitable Articles For Use With This Water Bath 

 

 

Plastic or glass laboratory test tubes that are sealed, waterproof and rated to withstand the intended temperature. 

 

Plastic or stainless steel laboratory test tube racks that are rated to withstand the intended temperature. 

 

An example of a suitable article would be a sealed plastic laboratory centrifuge tube filled with blood. 

 

Unsuitable Articles For Use With This Water Bath 

 

Any item where the article is not waterproof or may result in its contents coming into contact with the water bath medium (water). 

 

Items weighing more than 3 kilograms. 

 

An example of an unsuitable article would be an exposed piece of animal or plant matter. 

 

Suitable Environments For The Water Bath 

 

The water bath is intended for use in a clean laboratory environment only where adequate ventilation, a good power supply and provisions 
for routine cleaning are available. The water bath should not be used outdoors or in dirty, dusty, humid or windy environments. The 
acceptable operating conditions are outlined further in this User Manual.

  

 

General Operation 

 

The water bath is filled with water following the safety guidelines contained within this User Manual. 

 

The water bath is plugged into an appropriate power source. It is powered by an alternating current power supply with protective 
earth and with the appropriate receptacle, rated voltage and frequency for the country of its intended use. Further details on power 
requirements are outlined in this User Manual. 

 

A suitable article (as defined above) is placed on the mesh platform. 

 

The water bath is operated via a front panel interface consisting of heating temperature control buttons plus digital display and an 
on/off switch located at the rear of the bath. These controls allow the operator to set a required water temperature. 

 

The water bath should be operated strictly in accordance with the Operating Procedures outlined further in this User Manual. 

 

Operator Responsibility – Safety Considerations 

 
When operated in strict accordance with this User Manual, plus routine cleaning and maintenance being carried out, the product shall provide safe 
operation for the operator. The operator should be aware of the following before installing and operating the product :

 

 
Conditions of Operation 
 
*Note:

 The term “operator” referred to in this User Manual  is the primary person who has been tasked to install, maintain and train in 

the usage of this equipment. Other personnel shall be referred to as “Users”. 

 

 

The operator shall be aware that the protection provided by the equipment may be impaired if the equipment is used with accessories 
not provided or recommended by the manufacturer, is modified in any way or is used in a manner not specified by the manufacturer.

 

 

The operator is responsible for ensuring all users of the product are qualified to do so, and are well versed in common safety 
concepts. The product should only be operated by an adult who has read and understood this User Manual provided in the appropriate 
language in its entirety.

  

 

Any user must be informed by the responsible operator of any potential hazards that may arise through the use of this equipment in 
the course of their work, including any local environmental hazards not related directly to the water bath. They should also be able to 
demonstrate that they understand any preventative safety measures in operation prior to operating the equipment. 

 

The operator shall agree to accept responsibility for the use of the equipment in accordance with this User Manual, and be fully aware 
that the equipment is designed for commercial use.  

 

It is assumed that the user and operator have had experience in a commercial environment, and had appropriate training in how to 
perform their work safely in accordance with any local operational health and safety regulations. The operator and all users should be 
well versed in local emergency procedures as per the workplace safety regulations in effect. 

Summary of Contents for WB 000D Series

Page 1: ...rsion 1 0 29 02 2016 English www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Australia User Manual pri...

Page 2: ...uct to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product into the supplied packagi...

Page 3: ...eral Operation 4 OPERATOR RESPONSIBILITY SAFETY CONSIDERATIONS 4 Conditions of Operation 4 Safety Labels And Markings 5 OPERATING PROCEDURES 6 Identification of Controls Functions 6 Safety Warnings 7...

Page 4: ...e water bath is operated via a front panel interface consisting of heating temperature control buttons plus digital display and an on off switch located at the rear of the bath These controls allow th...

Page 5: ...s or questions relating to operator or user safety please contact the appropriate Ratek Service department before installing and operating the unit Contact details are provided in this User Manual Saf...

Page 6: ...ing a control or component of the water bath will be used to identify it The figure below indicates all key controls and components of the water bath with their corresponding numerical element labeled...

Page 7: ...place safety regulations in effect CAUTION If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired CAUTION The equipment must on...

Page 8: ...emptying NEVER operate the equipment if you believe it is damaged in any way NEVER operate the water bath if the mains power supply cable is damaged in any way NEVER use any sharp or metal objects ne...

Page 9: ...into the water bath ALWAYS use plastic stainless steel or plastic powder coated tubes racks and accessories in the water bath This will assist to prevent corrosion and thus help to extend the useful l...

Page 10: ...ated to match the power requirement as identified on the product identification label on the rear panel A surge protected power outlet is strongly recommended as it provides some protection for the eq...

Page 11: ...if contact with the element is made Setting the Required Water Temperature The water heating control circuit will retain the last used temperature set point set on the control panel so for many appli...

Page 12: ...y glove when removing the lid as the lid may be hot enough to cause burns or other injury NEVER remove the lid with your face directly over the water bath Steam may be present and burns or injury coul...

Page 13: ...placed below the level of the water bath on a firm hard flat surface If emptying into a sink or drain ensure you have clear and unobstructed access to it If the bath is fitted with a drain tap models...

Page 14: ...ferring to the section in this User Manual labeled Filling The Water Tank and allow 1 minute for the bath to cool The water bath can now be re operated as per the section of this User Manual labeled S...

Page 15: ...mple substances that may compromise the integrity or function of electrical insulation electrical components stainless steel components or water seals Ratek and the manufacturer are consulted prior to...

Page 16: ...ous injury or risk to life Refer the equipment to an authorized service technician for repair Over Temperature Safety Cutout Test The over temperature safety cutout should be tested periodically to en...

Page 17: ...to the up position marked with a circle and refill the water tank with 1 000mL of cold water to cool the element by referring to the section in this User Manual labeled Filling The Water Tank This wil...

Page 18: ...to 2 000 Metres Temperature 5 Celsius to 40 Celsius Ambient temperature will limit the minimum achievable water temperature Maximum relative humidity 80 for temperatures up to 31 Celsius decreasing l...

Page 19: ...ng The Water Bath provides a simple to operate user interface when used in conjunction with this User Manual If at any stage you experience abnormal operation anything other than that described in thi...

Page 20: ...n approved repairer or Ratek Instruments at the discretion of Ratek Instruments The limit of liability shall extend to the repair of the product only all other compensation claims are excluded from th...

Page 21: ...he unit to ensure safety for the service technicians Pack the unit into it s original packaging with the supplied mains power lead and use all original protective inserts If the original packaging is...