Preparing the Analyzer for Use
R&S
®
ZVA80
19
Quick Start Guide 1312.6808.62 ─ 02
Table 2-1: RF connection for left external test set
R&S ZVA80 BU connector
Ext. test set connector
Cable
Used at frequencies
PORT 1
–
SOURCE OUT
Diplexer
–
RF IN
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW / HIGH
PORT 1
–
REF IN
Diplexer
–
REF OUT
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW
PORT 1
–
MEAS IN
Diplexer
–
MEAS OUT
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW
PORT 3
–
SOURCE OUT
Converter
–
LO IN
R&S ZV-Z193, 3.5
mm
(M)
–
3.5
mm
(M),
1.5
m
HIGH
PORT 3
–
REF IN
Converter
–
REF OUT
CABLE EXT MEAS, 1.55
m
HIGH
PORT 3
–
MEAS IN
Converter
–
MEAS OUT
CABLE EXT REF, 1.55
m
HIGH
Table 2-2: RF connection for right external test set
R&S ZVA80 BU connector
Ext. test set connector
Cable
Used at frequencies
PORT 2
–
SOURCE OUT
Diplexer
–
RF IN
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW / HIGH
PORT 2
–
REF IN
Diplexer
–
REF OUT
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW
PORT 2
–
MEAS IN
Diplexer
–
MEAS OUT
R&S ZV-Z196, 1.85
mm
(M)
–
1.85
mm
(M),
0.6
m
LOW
PORT 4
–
SOURCE OUT
Converter
–
LO IN
R&S ZV-Z193, 3.5
mm
(M)
–
3.5
mm
(M),
1.5
m
HIGH
PORT 4
–
REF IN
Converter
–
REF OUT
CABLE EXT MEAS, 1.55
m
HIGH
PORT 4
–
MEAS IN
Converter
–
MEAS OUT
CABLE EXT REF, 1.55
m
HIGH
Fig. 2-7: Complete test setup for 2-port transmission measurement
Putting the Analyzer into Operation