![background image](http://html1.mh-extra.com/html/rands/zph/zph_getting-started_3338711031.webp)
Instrument Tour
R&S
®
Cable Rider ZPH
24
Getting Started 1321.0944.02 ─ 06
Risk of electrical shock
To avoid electrical shock, the DC input voltage must never exceed the value
specified on the housing.
Risk of instrument damage
●
Make sure not to overload the RF input and keep within the maximum
allowed signal levels. Refer to the datasheet for the maximum allowed
signal levels.
●
To avoid damage to the coupling capacitor, input attenuator or the mixer,
the DC input voltage of 50 V must never be exceeded.
4.3
Left View
1 = DC input
2 = Kensington lock slot
DC input
The R&S Cable Rider ZPH is supplied with power by the AC/DC transformer
power supply via the DC connector. You can also use the DC connector to charge
the battery.
Left View