Instrument Function
R&S
®
SMJ100A
265
Operating Manual 1403.7458.32 ─ 14
"Extern Step"
Generates the sweep signal step-by-step, manually triggered.
To trigger a sweep step, apply an external trigger signal. The step
width corresponds to the step width set for the rotary knob.
Example:
SOUR:SWE:FREQ:MODE STEP
SOUR:SWE:FREQ:SPAC LIN
SOUR:SWE:FREQ:STEP:LIN 1MHz
TRIG:FSW:SOUR EXT
SOUR:FREQ:MODE SWE
(External trigger)
"Extern Start/Stop"
Generates a continuously repeating sweep signal that is started,
stopped and restarted by subseqeunt external trigger events. The
sweep steps are performed automatically, controlled by the dwell
time.
Refer to the description of the rear panel for information on the con-
nectors for external trigger signal input (see
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR EAUT
SOUR:FREQ:MODE SWE
(External trigger)
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
Execute Single Sweep - Frequency Sweep
Starts a sweep manually. This trigger button is displayed in "Single" mode.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Reset Sweep - Frequency Sweep
Resets the sweep.
With the next trigger event, the sweep starts with at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL]
Start Freq - Frequency Sweep
Sets the start frequency.
Remote command:
RF Signal and Analog Modulations - RF Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...