Performing T-DMB/DAB Signal Generation Tasks
R&S
®
SMCVB-K156
45
User Manual 1179.1029.02 ─ 02
2. Optionally, activate "Stuffing" to adjust the TS data rate.
3. Specify IP TS channel properties, e.g. for "IP TS Channel > 1":
a) Define the input type, e.g. "Type > Multicast".
b) Specify the "Multicast Address", that is the destination IPv4 address of the IP
connection.
Note:
Use the destination address also in the IP data source, e.g. a stream
program.
c) Specify the port, that is the destination port of the IP connection.
Note:
Use the destination port also in the IP data source.
d) Specify the
source address, that is the source IPv4 address of the IP
connection.
A source address different from "0.0.0.0" accepts only data originating from the
specified IP address.
e) Optionally, to check availability of the "IGMPv3 Source Address", click "Ping
Source Address"
● If "Ping Result > Ping: Successful", the source address is available.
● If "Ping Result > Ping: Transmit Failed. Destination Host Unreachable", try
another address.
f) Optionally, specify a name for the IP connection, e.g. "Alias > Service".
g) Activate the IP channel, select "Input IP > On".
The R&S
SMCV100B is prepared for receiving IP TS data, that is input at the "IP
Data" connector.
5.1.2
How to Apply an External TS Input Signal
To connect to the external TS input interface
1. Use a double-shielded 75
Ω
BNC cable for connection between R&S
SMCV100B
See also Section "Cable selection and electromagnetic interference (EMI)" in the
R&S
SMCV100B Getting Started user manual.
Configuring the Input Signal