Instrument configuration
R&S
®
QAT100
28
User Manual 1179.3073.02 ─ 02
6
Instrument configuration
The R&S
QAT provides various settings for the test setups described in
on page 20. You can access all settings via the controls on
the rear panel of the R&S
QAT.
Configuring a LAN connection
When you use the R&S
QAT in a LAN, configure the network properties.
1. Enter "Network" settings.
2. Turn "DHCP" (dynamic IP address assignment) on or off.
When dynamic IP address assignment is on, the network automatically assigns a
free IP address to the connected R&S
QAT.
3. When you turn off DHCP, configure the IP address (IPv4), subnet mask and gate-
way manually using the corresponding
. Refer to your system
administrator for information about network properties you can use.
Configuring the frequency reference
Some test setups use an external or the internal frequency reference to synchronize
the frequency between the devices in the test setups.
1. If you are using an external frequency reference to synchronize the frequency:
a) Select "QAT Settings" > "Reference" > "External".
b) Select the frequency you want to use. Supported frequencies are 10
MHz and
100
MHz.
If you are using an R&S
AREG800 as the source of the frequency reference,
you can also use 3.2
GHz.
2. If you are using the internal frequency reference of the R&S
QAT to synchronize
the frequency:
a) Select "QAT Settings" > "Reference" > "Internal".
b) Select the frequency you want to use. Supported frequencies are 10
MHz and
100
MHz.
Selecting the radar band
1. Select "QAT Settings".
2. Select "Radar Band".
3. Select one of the available radar bands.
● 76 GHz to 77 GHz
● 77 GHz to 81 GHz
● 76 GHz to 80 GHz