Remote Control Commands
R&S
®
NRX
236
User Manual 1178.5566.02 ─ 07
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:INFO?
[<ITEM>]
Effective for continuous average measurements.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Query parameters:
<ITEM>
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Effective for continuous average measurements.
Sets the size of the result buffer.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<count>
Range:
1 to 131072
*RST:
1
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average measurement. If enabled, all
results generated by trigger events are collected until the buffer is filled. Thus, a higher
data rate is achieved.
You can set the size of the buffer with
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100