Remote Control Commands
R&S
®
NRX
235
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]POWer:REFerence
<ref>
Sets the reference value for the relative power indication.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<ref>
Range:
-290.0 to +110.0
*RST:
+30.0
Default unit: dBm
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
on page 55
See
13.5.6
Continuous Average
Further information:
●
Chapter 7.1, "Continuous Average"
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
.............................................................235
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
..........................................................235
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:INFO?
.............................................................236
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
...............................................................236
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
.............................................................236
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
Effective for continuous average measurements.
Clears the contents of the result buffer.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Effective for continuous average measurements.
Queries the number of results that are currently stored in the result buffer.
Measurement Settings and Results