Remote Control Commands
R&S
®
NRX
343
User Manual 1178.5566.02 ─ 05
OUTPut:TRIGger:SOURce
<source>
TOUT
is set.
Sets the trigger source.
Parameters:
<source>
SENS1 | SENS2 | SENS3 | SENS4 | EXTernal | CHKSource
*RST:
EXTernal
Manual operation:
See
"Trigger Source for Trigger Output"
14.9
Zeroing
Further Information:
●
........................................................................................... 343
..................................................................................343
CALibration<undef>:ALL:ZERO:AUTO
............................................................................. 344
CALibration<Sensor>:ZERO:FAST:AUTO
.........................................................................344
CALibration<undef>:ALL:ZERO:FAST:AUTO
.................................................................... 344
CALibration<Sensor>:ZERO
Performs zeroing for the power sensor connected to selected port.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout. Use
to recognize the end of a zeroing procedure.
Example:
CAL2:ZERO
Usage:
Event
CALibration<Sensor>:ZERO:AUTO
<auto>
Performs zeroing using the signal at the power sensor input.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout. Use
to recognize the end of a zeroing procedure.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Zeroing