Remote Control Commands
R&S
®
NRX
285
User Manual 1178.5566.02 ─ 05
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
14.5.2.3
Burst Average Calculation Functions
Parameter list
For the calculation functions of the burst average measurement, the following parame-
ters are used.
●
<dtolerance>
Mandatory. Length of a time interval during that the power level can drop below the
trigger level without being interpreted as end of the power pulse.
Corresponds to
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
.
●
<start_exclude>
Mandatory. Amount of time at the beginning of a timeslot or integration period that
is not evaluated.
Corresponds to
[SENSe<Sensor>:]TIMing:EXCLude:STARt
.
●
<end_exclude>
Mandatory. Amount of time at the end of a timeslot or integration period that is not
evaluated.
Corresponds to
[SENSe<Sensor>:]TIMing:EXCLude:STOP
.
●
<expected_value>
Optional. Value that is expected for the measurement.
●
<resolution>
Optional. Limit up to which the measurement result should be free of noise.
Corresponds to
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.
●
<source_list>
Optional. Primary and secondary sensor. The number for each sensor is preceded
by the character
@
. The entire expression is enclosed in parentheses.
Example: (
@
3),(
@
2)
Sensor C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>,
<start_exclude>, <end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
FETCh<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>, <start_exclude>,
<end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
Calculation Functions