Remote Control Commands
R&S
®
NRPxxS(N)
127
User Manual 1177.5079.02 ─ 10
After zero calibration, query the static error queue (
responses are possible:
●
0
No error, the zero calibration was successful.
●
-240
Warning, zero calibration failed. See also the example.
Suffix:
<Channel>
.
1...4
Measurement channel if more than one channel is available.
Parameters:
<state>
ONCE
Only valid parameter for this command.
0
Return value if no calibration is in progress.
*RST:
OFF
Example:
*CLS
CAL1:ZERO:AUTO ONCE
Performs zeroing. Takes several seconds.
SYST:SERR?
Query
-240
Response: Warning; Zero Calibration failed; Results Degrading.
Manual operation:
See
9.10
Testing the Power Sensor
The selftest allows a test of the internal circuitry of the sensor.
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps
Offset Voltages
and/or
Noise Voltages
.
TEST:SENSor?
[<Item>]
Triggers a selftest of the sensor. In contrast to
, this command returns detailed
information that you can use for troubleshooting.
Query parameters:
<Item>
String
Usage:
Query only
Manual operation:
See
Testing the Power Sensor