Remote Control Commands
R&S
®
NGL200
89
User Manual 1178.8736.02 ─ 02.01
// selects the output channel, sets the step width
// and increases the voltage in the selected channel
// by the set 4 Volts
VOLT DOWN
// decreases the voltage in the selected channel
// by the set 4 Volts
VOLT:STEP?
// queries the voltage step size
// repsonse: "4.000"
..................................................................................................89
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer
....................................................... 89
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]
...................................................... 90
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
.......................................................... 90
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
................................................91
[SOURce:]ALIMit[:STATe]
<state>
[SOURce:]ALIMit[:STATe]?
Activates/Deactivates or Queries the safety limit for voltage.
Parameters:
<state>
1 | 0
1
Activates the safety limit for voltage.
0
Deactivates the safety limit for voltage.
Example:
See
Example "Configuring the output voltage"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer?
Sets/Queries the lower safety limit for voltage.
Setting parameters:
<voltage>
numeric | MIN | MAX
numeric
Numeric value for safety limit.
MIN
Min value for lower safety limit.
MAX
Max value for lower safety limit.
Range:
0.000E+00 to 2.000E+01
Increment: 0.001
*RST:
0.000E+00
Example:
See
Example "Configuring the output voltage"
Configuration Commands