Remote Commands
R&S
®
FSW
954
User Manual 1173.9411.02 ─ 43
14.5.8.2
Configuring a Sweep List
The following commands configure the sweep list for spurious emission measure-
ments.
The sweep list cannot be configured using remote commands during an on-going
sweep operation.
Useful commands for configuring the sweep described elsewhere:
●
●
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
●
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
Remote commands exclusive to spurious measurements:
[SENSe:]LIST:RANGe<ri>:BANDwidth:RESolution
............................................................ 954
[SENSe:]LIST:RANGe<ri>:BANDwidth:VIDeo
................................................................... 955
....................................................................................955
..................................................................................955
................................................................................... 956
[SENSe:]LIST:RANGe<ri>:DETector
.................................................................................956
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE
............................................................................956
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation
................................................................... 957
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO
......................................................... 957
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe
...................................................................958
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
.................................................................958
[SENSe:]LIST:RANGe<ri>:LIMit:STARt
.............................................................................959
[SENSe:]LIST:RANGe<ri>:LIMit:STATe
.............................................................................959
[SENSe:]LIST:RANGe<ri>:LIMit:STOP
............................................................................. 959
[SENSe:]LIST:RANGe<ri>:POINts[:VALue]
....................................................................... 960
................................................................................... 960
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
.......................................................................... 960
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO
................................................................ 961
[SENSe:]LIST:RANGe<ri>:TRANsducer
........................................................................... 961
[SENSe:]LIST:RANGe<ri>:BANDwidth:RESolution
<RBW>
This command defines the resolution bandwidth for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<RBW>
Resolution bandwidth.
Refer to the data sheet for available resolution bandwidths.
Default unit: Hz
Example:
LIST:RANG2:BAND:RES 3KHZ
Configuring and Performing Measurements