Remote Commands
R&S
®
FSW
924
User Manual 1173.9411.02 ─ 43
<ri>
1...30
Selects the measurement range.
Parameters:
<FilterType>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FSW-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FSW-K54) option
P5
5 Pole filters
*RST:
NORM
Refer to the datasheet for available filter bandwidths.
Manual operation:
See
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FSW will not measure the
ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FSW will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span
For more information see
Chapter 7.7, "Spurious Emissions Measurement"
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Example:
LIST:RANG2:STAR 2MHZ
Manual operation:
See
See
Configuring and Performing Measurements